SPESIFIKASI KEDOKTERAN,KESEHATAN DAN KB.

PENGADAAN ALAT KEDOKTERAN, KESEHATAN DAN KB
RSUD H. ABDUL MANAP KOTA JAMBI
DANA APBN P TAHUN ANGGARAN 2011

SPESIFIKASI

NO.

NAMA BARANG

VOLUME

PERALATAN LABORATORIUM
1

BLOOD BANK REFRIGERATOR

1 Unit

Specifications :


Exterior (Cabinet) : Material: Galvanized steel plate, Powder coating
Interior (Chamber) : Vol: 245 l, Material: Galvanized steel plate, Powder coating
Door : Double anti-fog glass
Insulation : Non-CFC Polyurethane foam or foam panel, Thichness: 70mm

Drawers : Material : Stainless steel plate
Compressing Unit : 1/5 HP hermetically sealed compressor x 1, Air-cooling type
Cooling Unit : Finned tube coil type evaporator with blower,
Automatic off-cycle defrost system
Refrigerant : R-134a, Charge: 280g
Operating Temperature :Approx. 1.0º C ~ 6.0º C
Power Source : AC 230V, 50/60Hz
Power Consumption : 450W
Chamber Lighting : Fluorescent lamp, 20W, 1ea
Temperature Recording : Recording range: -10ºC ~ +40ºC , 7-day Revolution,
Ø125 mm Diagram, 1.5V (LR6) Battery driven, Fiber tip pen u
2

SEROLOGICAL CENTRIFUGE


1 Unit

Specifications :

Maximum speed
Maximum RCF
Control system
Speed set range
Speed set step
Timer set range
Timer set step
Motor
Supply values
Power consumption
3

: Range 4500 - 5000 rpm
: 2822 X Gram (g)
: programmable microprocessor control
: 1000 – 5000 rpm

: 10 rpm
: 1-99 minutes and hold position
: 1 minute
: induction motor
: 230 V 50/60 Hz, 110 V, 60 Hz
: 350 W

Dry Heat Sterilizer & Incubator 32 LT
Specifications :

Cabinet : Stainless steel
Chamber : Stainless steel/ Volume 32 L
Insulation : Mineral Wool, t 50 mm
Power Source : AC 230V, 50/60 Hz
Power Consumption : 1.200W

2 Unit

Mode : Sterilizer / Incubation
Temperature Control : Microprocessor, PID controller

Convection Method : Natural Convection
Temperature Range :
- Sterilization Mode: 140°C, 160°C, 180°C (available user cycle 35°C~220°C)
- Incubatio Mode : 37°C, 42°C, 52°C, 60°C, 70°C (available user cycle 35°C~80°C)

Display : Graphic Display, LCD module
Environment Conditions
- Atitudes up to 2.000 meters
- Temperature range of +5°C ~ +40°C.
- Maximum reative humadity of 80% for temperatures
up to 31°C, decressind linearly to 50% at 40°C
- Main supply voltage fluctuation of ±10% of nominal
- Installation category (overvoltage category) II,
Pollution Degree 2
4

MICROSCOPE BINOCULAR

3 Unit


Specifications :
- Binocular head
- Widefield ocular 10x, Plan field ocular 16x
- Standard achromatic objectives (4x, 10x, 40x and 100x oil)
- X-Y stage with vernier scales
- Coaxial fine and coarse adjusment
- 1.25 N.A. Abbe condenser with adjustable iris diaphragm
- 6V/20W halogen illumination light system with intensity control
5

PLASMA EXTRACTOR

1 Unit

Specifications :
BENEFITS
* Labor saving
* Perfect for automation of componens separation
* Enables operator to multi-task
* Up to 5 electronic expressors can operate on one A/C power adaptor

* Eliminates use of multiple power adaptors
* 1-year warranty
Sensor type : Infrared interruption sensor
Clamping power source : Geared motor
Clamp speed : ~ 1 secons from detecting red cells
Controls : Power, open, close, start, sencitivity
POWER REQUIREMENTS
Power source : A/C power adaptor
(A/C 100-120V or 220-240V A/C in, 18 V out
Power consumption : 10 Watts max
ACCESSORIES INCLUDED
Power adaptor
Jumper cable (for tandem connections)
User manual
6

HAND SEALER

1 Unit


Specifications :
Base Material : Alumunium Alloy & SUS
Rollers : Special Composite Steel, Non-Rust
Cutters : Not Available
Features : Self-Centering Rollers
Cutter : None
7

1 Unit

HEMOSCALE
Specifications :
Collection Volume : 0 - 1199 mL
Weight Display : 0 - 1199 Grams
Elapsed Time Display : 0 - 1199 Grams
Target Volume range : 0 - 1199 mL Default Value 450 mL
or 500 mL with 5 mL Increments
Accuracy
Weight Tolerance : +/- 1 grams
Weight Accuracy : 0 grams to + 3 grams

Time Display Accuracy : 1 second maximum
LCD Custom Designed Display : 110V - 120V A/C, 220V - 240V A/C, 50 - 60 Hz
Power Requirements : Rechargeable Ni-Cad 12V DC
Battery Pack : Rechargeable Ni-Cad 12V DC
Accessories Include :

8

- Ni-Cad Battery Pack
- Battery Recharger unit
- A/C Power Adapter
- User Manual
- Carrying/Transport Case
- Mixer Stand

HEMOGLOBINE METER

1 Unit

Specifications :

Measuring Range : 0 -240 g/L
Tolerance : < 2 %
Linear Range : < 5 % FS
Dilution Ratio : 1:251
Testing Time : < 4 s (maximum absorption)
Stability : < 1 g/L/h (25 ± 10ºC)
Continuous working time : > 10 hours
Sample absorption : 1 - 9 mL, ten grades adjustable
Standard setting : 0 - 240 g/L adjustable (whole number times of ten)
Electrical : AC220 V ± 10% (50Hz)
Power Consumption : < 10 W
9

COOL BOX
Specifications :
Capacity : 25 liters
Power supply : AC 230V/40W & DC 12-24V/40W
Temperature Range : -18ºC to + 10ºC

2 Unit


Ambient Temperature : -10ºC up to + 40ºC
Power consumption : 40 watt
10

1 Unit

IMUNOLOGI ANALIZER + UPS
Specifications :
System
Automatic rotary-random access panel input
Batch mode
Hingga 30 tes berbeda tiap running (1 jam 20 menit)
Metoda
ELISA
Compliment Fixation Test (CFT)
Sampel
Serum
Display
Touch-screen LCD 6” STN color display

Printer
Dot thermal, 832 dot/line
Ukuran kertas 112 mm continuous roll
Panjang Gelombang
450, 640, 650 nm
Packaging
36 atau 12 devices
The single test device contains all the reagents necessary
Master curve procedure with 1 point calibrator included in the kit
Internal CPU
16 bit Microprocessor
Multi I/O Ports
1 Mb internal permanent memory
512 Kb RAM
Kondisi Operasional
Range temperature
: 18-300C (64-860F)
Kelembaban
: < 80%
Ketinggian dari dasar
: < 2000 mt (6.562 kaki)
Cahaya
: Jangan berada pada tingkat cahaya yang sangat terang
Hindari debu dan gas berbahaya
Keperluan Listrik
Arus listrik
Line voltage
Absorbed power
Level kebisingan
Fuses
Instrument & Accessories
Washing Buffer

:
:
:
:
:

105< >240 VAc ; 50-60 Hz
110/220 Volt