LPSE POM spek adendum

Lampiran Addendum Spesifikasi Teknis
Dok. 01.02.100.04.12.0678
PENGADAAN ALAT LABORATORIUM BALAI BESAR POM di BANJARMASIN
RINCIAN SPESIFIKASI TEKNIS PENGADAAN ALAT TAHUN 2012

No.

SPESIFIKASI

2 HPLC with auto & manual injector, UV-VIS Detector with manual loop 500µl *)
Specifications :
- System Pump
: Binary and Quaternary
- Flow rate setting range
: 0.001mL/min. to 10 mL/min.
- Flow rate accuray
: ≤ %
- Flow rate precission
: ≤ . %
: minimal 4
- Number of solvent mixed

- Mixing - concentration accuracy
: ≤ . %
Detector uv-vis
- Light source
: D2 lamp
- Wavelength range
: 190-700 nm or wider
- Wavelength accuracy
: ± 1 nm
- Wavelength precision
: ± 0,1 nm
- Function
: Dual Wavelenght
- Linearity
: 2.5 AU
Column Oven
- Temperature setting range
: 4°C to 80°C or wider
- Temperatur control precision
: ≤ , oC

- Coulumn storage
: min 4 columns
Autosampler
- Injection Precision
: < 0,5% or better
- Injection volume
: 0,1-100 uL
- Number of sample vials
: minimal 100 vial@±1-2 ml
- Sample carryover
: min 0.01%
Manual injector

SATUAN

JUMLAH

Unit

1


Can inject from 1 µl to 5 ml with high accuracy and precision
System Controller
Number of connectable units
: min 8 units

Computer
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor
LCD 18,50", Laser colour printer, USB 2 port
Data Analysis & Reporting for food and drug
Include :
a.
b.
c.
d.
e.
f.
g.
h.

i.
j.
k.
l
m.
n.
o.
p.
q.
r.
s.

Maintenance Kit for Pump
Auto rinsing kit for Pump
Mixer SUS
Sample vial, 1 - 2ml (100pc/ pk) min 2 pkg
Micro syringe 100 uL 1 pcs
C18 Coulumn 4.6x250 mm, 5 um min 2 pcs
C 8 Coulumn 4.6x250 mm, 5 um min 2 pcs
C18 Coulumn 4.6x150 mm, 5 um min 1 pcs

Check valve inlet min 2 pcs
HUB W/ Cable
Tools Kit
UPS 5 KVA
Reservoir Tray
Sample loop 500µl for manual ijnector
Instalasi
Training
Original manual book
Factory cerificate
Guarantee min 1 year

3 HPLC with autosampler, UV-VIS Detector *)
Specifications :

Unit

2

Pompa

- System Pump
- Flow rate setting range
- Flow rate accuray
- Flow rate precission
- Gradient for easy method development
- Number of solvent mixed
- Mixing - concentration accuracy
Detector UV-Vis
- Light source
- Wavelength range
- Wavelength accuracy
- Wavelength precision
- Flow cell design

- Function
- Linearity
- Lamp optimization

Column Oven
- Temperature setting range

- Coulumn storage
Autosampler
- Injection volume
- Number of sample vials
- Sample carryover
- Injection Precision
Computer

:
:
:
:
:
:
:

Quaternary
0.000 mL/min. to 10.000 mL/min. (0.050 to 5.000 mL/ min typical)
≤ %
≤ . %

min 10 gradient curves function
minimal 4
≤ . %

:
:
:
:
:

D2 lamp
190-700 nm or wider
± 1 nm
± 0,1 nm
maximizes energy throughtput, eliminates refractive index effect,
ensuring the highest sensitivity while maintaining optimal spectral
performance
: Dual Wavelenght
: 2.5 AU
: Designed to automatically maximize signal to noise in both visible

and UV spectrum with the use of the single D2 lamp, extends the
useful life of the lamp for consisten result overtime
: 20°C to 65°C or wider
: min 2 columns
: 0,1-100 uL
: minimal 100 vial@±1-2 ml
: min 0.1%
: < 0,5% or better

Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor
LCD 18,50", Laser colour printer, USB 2 port
Data Analysis & Reporting for food and drug analysis.
Include :
a.
b.
c.
d.
e.
f

g.
h.
i.
j.
k.

Tool kit
C18 Coulumn 4.6x250 mm, 5 um min 2 pcs
C18 Guard Coulumn 4.6x20 mm, 5 um min 2 pcs
Sample/standard filtration, Ø 13 mm 0.45 µm
Sample vial, 1 - 2ml (100pc/ pk) min 1 pkg
UPS 2 KVA
Instalasi
Training
Original manual book and software
Factory cerificate
Guarantee min 1 year

4 Aquadestillata (Glass-Destilation) *)
Spesification

a.
Capacity
b.
Cooling water requirement
c.
Conductivity
d.
Evaporator material
e.
Condensor material
f.
Heating element material
g.
Low water cut off
h.
Energy saving
i.
Distillate withdrawl
j.
Cooling water outlet
k.
Impurity detector
l.
Housing
m.
Cooling water pressure

unit
:
:
:
:
:
:
:
:
:
:
:
:
:

min. 8 L/ hour
± 144 L/ hour
± 2.2 µs/cm at 20°C
borosilicate glass
borosilicate glass
Quartz sheathing
in condensor, auto power cut off
through distillation of heated cooling water
Right hand side
Ø 12.7 mm
Automatic
Galvanized steel, powder coated with epoxy resin
> 3-7 bar

1

n
o.
p.
q.

Included of
-

Dimension (WxDxH)
Power input
Electrical consumption
Another features

:
:
:
:

max 650 x 365 x 390 mm
400 V/ 50-60 Hz
max. 6.0 kW
- Electronic impurity detector by automatic water change
- Water saving through automatic water cut-off
- Sterillisation of the condensers by steam
- CO2 degassing through outlet in the condenser
- Water supply and cooling water outlet with hose
connection 1/2 inch.

Wall or bench mounting
Instalation
Guarantee min 1 year
Original operational manual book

5 Aqua reverse Osmosis System *)
Spesification
a.
Nominal permeate flow
b.
Product water volume
b.
Inlet feed water
c.
Ionic content rejection
d.
Organics, particulate
microorganism rejection
e.
Dimension (HxWxD)
f.
Operating weight
g.
Built in reservoir volume
h.
Electrical feed voltage
i.
Tap (feed) water connection
j.
Tap (feed) water pressure
k
Water quality
Accessories/catridge used
RO membrane catridge (2 pcs)
Sanitization chlorine tablet (2 pack)

unit
:
:
:
:
:

min 16 L/h
160 to 320 L/day
tap water
> 96 %
> 99 %

:
:
:
:
:
:
:

max. 45.7 x 25.5 x 31.5 cm
max 15 kg
6L
220 - 240 V
1/2" Gaz M
0.5 - 6 bar
type 3

1

Included of
-

Vent filter for PE reservoir (2 pcs)
Pretreatment kit (2 pcs)
100 L PE reservoir
(1 pc)
1200 L tank
(1 pc)
Water pump (1 pc)
Installation
Guarantee min. 1 year
Original operational manual book

10 Gas Chromatography (GC) with detector FID, ECD *)
Specifications :
Column Oven
- Temperature range
- Cool down rate
- Programmability
- Retention time repeatibility
Injection Port
a. Split/Splitless Injection Unit
( 2 set )
- Temperature range
- Pressure setting range
- Split ratio setting range
- Total Flow setting range
- Programmable steps
b. Injection Unit for Widebore Column
Temperature range
Pressure setting range
Total Flow setting range
Programmable steps
Detectors

Unit

:
:
:
:

+ 4o to
⁰C
˂
i utes fro
°C to
Up to 20
≤ ,
% for -deca e

:
:
:
:
:

Room temperature +10 o C to 400 o C or better
0 - 900 kPa or better
1-8000 or better
0 -1200 mL/min or better
7

:
:
:
:

Room temperature +10°C to 400°C or better
0 to 900 kPa or better
0 to 1200 mL/min or better
7

°C

1

Four controlledr detectors can be installed simulatenously and individually temperature
- Flame Ionization Detector (FID)
Specification :
a.
Temperature range
: to 400°C or higher
b.
Minimum detection quantity
: 1.5 pgC/s
c.
Dynamic range
: 107 or greater
- Electron Capture Detector (ECD)
Specification :
a.
Temperature range
b.
Minimum detection quantity
c.
Dynamic range
Autoinjector / Autosampler
a.
Sample Injection Methode
b.
Number of Samples
c.
Number of sample injections

: to 350°C or higher
: 6 fg/s
: 104 or greater

: Liquid sample injection via special Microsyringe.
: Minimum 100 vial @1 - 2 mL
: 1 - 99 injections per sample.

Computer
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor
LCD 18,50", Laser colour printer, USB 2 port
Data Processing and Data Analysis
Accessories :
-

Carrier Gas Supply pipe min 2 pcs
Gas filter, molecular sieve
He Gas + bottle
Pressure Regulator He
H2 Gas Supply Pipe
H2 Gas + bottle
Pressure Regulator H2
Oil-less air compressor

-

Regulator filter
Air filter for air compressor (silica gel)
Joint Adapter
Air supply pipe, 2 pcs
10-µL microsyringe min 2 set
Capillary Column Q-Plot, 30m, 0.53mm ID, 0.25um min 1 pcs
Capillary Column DB-1, 30m, 0.25mm ID, 0.25um min 1 pcs
Capillary Column Stabilwax, 30m, 0.25mm ID, 0.25um min 1 pcs
Capillary Column DB-5, 30m, 0.25mm ID, 0.25um min 1 pcs
Capillary Cutter
ECD Control Unit
ECD Detector Unit
N2 Gas + bottle for ECD
Presure Regulator N2 for ECD
Oxygen Trap for ECD
AUX Temperature Control Unit
RS-232 Cable
1 - 2 ml Sample Vial Set with white cap & septum, 100 pcs/set min 5 set
UPS 5 KVA
Instalation
Training
Guarantee min. 1 year
Original operation manual
Factory Certificate
Tool Kit
Training to BAPETEN to get ECD license

16 Trinocular Microscope
A rugged time tested design with true Kohler ilumination and high quality optics, suitable for day to day use in the laboratory. The
options below are combined into twelve standard configurations to suit a wide range of biological applications and budgets
Spesifications :
- Viewing Heads

Unit

1

Available with standard binocular heads or trinocular heads for photomicrography, 30 o inclined eyetubes with adjustable
diopter settings and interpupilarrydistance, 360 o rotatable
- Eyepieces
Paired DIN KHW 10x, FN
- Objective Changer
Smooth operating, ball bearing mounted, quadruple nosepiece
- Objectives
S. Flat Achromats, SM Planachromats, Planachromats, or phase contrast
- Stage
Attached mechanical stage : 120 mm x 140 mm surface with 75 mm (X) by 40 mm (Y) movement, graduated with vernier
readings to 0.1 mm, drop down right or left handed controls
Flat top stage : 171 mm x 140 mm surface with 80 mm (X) by 52 mm (Y) movement, graduated with vernier readings to 0.1
mm, detachable speciment quide, drop down right of handed controls.
- Condensers
Standard models have either an Achromatic NA 1.25 condenser with iris diaphragm and filter tray for brighfiel work, simple
NA 1.25 slider phase condenser, or an NA 1.25 Zernike phase condenser with filter tray and independently centerable annuli
for phase contrast. Cardioid type NA 1.25 darkfield condenser for high power ( 100x objective only) darkfield work is optional
- Illuminator
Built in 6 V 300 W, variable intensity Koehler illumination for apertures up to 1.25
Head
: Trinoculer
Eyepieces
: KHW 10 x FN 20
Objectives
: SM Plan Brightfield 4 x, 10 x, S40 x, S 100 x oil
Mech stages
: Attached
Condensers
: Achromatic NA 1.25 with blue filter
Illumination
: Koehler 6 V 30 W halogen
Accessories
- USB 2.0 Camera 1.5 megapixel camera
High speed USB 2.0 interface eliminates a framegrabber and facilitates ease of instalation on both laptop and dekstop
computers. Infinity Capture software application included.
Max compatible, on board-image processing of the infinit lite's progressive scan 1.5 megapixel image sensor result in
accurate color fidelity
Samples : 15 fps at full 1440x1080 resolution and 60 fds at 640x480

- Mount adapter with 0.5 x lens
- Personal computer
Dekstop PC, processor intel core i-5, memory 4 GB DDR3, VGA 1 GB, SATA 5400 rpm, DVDRW, Gigabit LAN 10/100/1000
Mbps, USB Keyboard,'Optical Mouse, Windows 7 home premium, Monitor LCD 20, Power supply 300 W, printer.
Included
- Guarantee 1 year
- Instalation
- Training
- Original operation Manual
21 Safety Cabinet (Yellow) For Flammable
Specification :
a.
Material

b.

Construction

c.
d.
e.

Load capacity
Complience with
Safety device

f.
g.
Included :
-

Dimensions : H x W x D
No. of shelves / door
Instalation
Guarantee min. 1 year

Unit
: Galvanized steel min. 51 mm, fully painted inside-outside with a
durable lead free epoxy/ polyester powder paint for chemical
resistance.
: Compact 3-point stainless steel bullet self-latching system and
welded construction
: 90 gallon/ 159 kg
: OSHA and NFPA standards, and most are FM approved
: - dual vents with flame arresters
- Shut and latch automatically when fusible link melt at 165°F
(74°C) under fire conditions
- double-walled steel with 38 mm of insulating air space for fire
resistance
- paddle handle with slip resistant
- fail safe closing mechanism.
: min 1651 x 1092 x 864 mm
: 2/2

2

-

Original operation manual

25 4.0 Vertical Auto-desiccator cabinets
Protecting valuable humidity-sensitive items, such as reagents. It's full automatic system. electronic dehumidification
system that automatically regenerates the desiccant and reduces the relative humidity to about 25% within 8 hours
A hygrometer dial in the door allows for easy monitoring of relative humidity.
Spesification
a.
Material
b.
Range of humidity control
c.
Power supply
d.
Internal Volume
e.
Dimension (HxWxD)
f.
Color
g.
Accessories

h.

:
:
:
:
:
:
:

Unit

1

Unit

3

Co-polyester plastic
up to 25% RH
230 V AC
1.9 cu feet
max 51.8 x 34.1 x 41.4 cm
Clear
- 2 latches and loops for lock or tamper evident seal
- 3 perforated shelves that are removable

Include
Instalation
Guarantee min. 1 year
Original operation manual

27 2.0 Auto desiccator cabinets
Protecting valuable humidity-sensitive items, such as reagents. It's full automatic system. electronic dehumidification
systemthat automatically regenerates the desiccant and reduces the relative humidity to about 25% within 8 hours
A hygrometer dial in the door allows for easy monitoring of relative humidity.
Spesification
a.
Material
: Co-polyester plastic
b.
Range of humidity control
: up to 25% RH
c.
Power supply
: 230 V AC
d.
Internal Volume
: 1.17 cu feet
e.
Dimension (HxWxD)
: 31.4 x 34.1 x 41.4 cm
f.
Color
: Clear

g.

Accessories

h.

Include
Instalation
Guarantee min. 1 year
Original operation manual

: - 2 latches and loops for lock or tamper evident seal
- 2 perforated shelves that are removable