LAMPIRAN ADENDUM SPEK TEKNIS ALAT BPOM PLKY
ADDENDUM SPESIFIKASI TEKNIS
PENGADAAN ALAT LABORATORIUM BPOM DI PALANGKARAYA
TAHUN ANGGARAN 2012
NAMA ALAT DAN SPESIFIKASI
NO
1 Analytical balance*)
Specification :
*.
Capacity
Max 250 g
*.
Reproducibility
0,2 mg
*.
Min piece wight at piece count
*.
*.
*.
*.
*.
Readabilty
Linearity
Overall dimensions (include draft shield)
Weighing plate
Includes
2 Fumehood *)
Instalation, Training, Guarantee min. a year, Original
operation Manual, Dust Cover.
*.
Face velocity 0,5 m.s
*.
*.
*.
*.
Light intensity
Construction Main Body
Power Supply
Accesories
Included of
Min 1720 X 680 X 1200 mm
580 mm, vertical rising sash
Max : 1795 cmh, Min : 900 cmh
780 lux
Phenolic resin Laminate or equivalent
220 - 240 V, 50-60 hz
Base Storage Cabinet with Standard Fittings for 6ft/;
Exhaust Fan/Blower Polypropylene
(diameter :
250mm)
Instalation, Guarantee min. a year, Original operation
Manual
3 HPLC with UV-Vis detector and autosampler*)
Specification :
*. PUMPS
*.
Solvent delivery method
Parallel-type double plunger
*.
Flow rate accuracy
No more than 1% or 0.5uL/min whichever is greater
(0.01 to 2m L/min).
No more than 0.06% or 0.02 min SD. whichever is
greater
*.
*.
Flow-rate setting range
Flow rate precission
Plunger capacity
Maximum discharge pressure
Typical Pulsation
UNIT
80 mm
± 270 mm ( 10.6")
*.
2
WDH : 210 x 330 x 335 mm
Exhaust outlet
Max sash opening
UNIT
0,6 mg/piece
*.
*.
1
Max 0.2 mg
Min 1800 x 875 x 1500 mm
Internal dims
SATUAN
UNIT
± 0.1 mg
Specification :
*. External dims ( LxDxH in mm )
*.
JML
2
2 set
0.0001mL/min. to 10.0000mL/min.
10 uL.
40MPa.
0.03 MPa (for water at 1.0 mL/min and 7 MPa).
Constant-pressure solvent delivery Supported.
*.
*.
Plunger rinsing mechanism
*.
Safety measures
OVEN COLUMN
*.
Temperature range
Manual rinsing or automatic rinsing using optional
product.
Liquid-leakage sensor, high pressure/low-pressure
limits.
4°C to 85°C or wider
2 set
*.
AUTO SAMPLER
*.
Injection method
*.
Injection-volume setting range
*.
Injection volume precision
*.
*.
*.
*.
*.
*.
*.
*.
Number of processed samples
175 (1ml vials)
Dimensions wight
Power requirements
Lamp
Wavelength range
WORKSTATION :
*.
Hardware
*. Software
ACCESSORIS
*.
Maintenance Kit
*.
Mixer
*.
*.
*.
*.
100 VAC, 100 VA, 50/60 Hz
D2 lamp
Intel Core i3-2100 Processor, 2GB 1x2G
CMel/CTino D3-1333, Windows 7 Professional 32bit, HDD 500GB 7200RPMSATA, Intel Graphics
Media Accelerator (GMA) X4500HD, HP USB
Optical Mouse , USB Standard Value Keyboard, HP
15-in-1 Media Card Reader, HP SuperMulti (JB) 1st
ALL, Monitor HP LCD 18,50", Laser Jet HP 1102,
speed up 17 ppm, 1200 dpi effective output, USB 2
port
Data analysis & reporting for food and drug analysis
2 set
2 set
2 set
4 set
4 set
4 set
2 set
2 set
2 set
2 set
2 set
ANALYTICAL COLUMN C8, 15CM, 5 um
HUB W/ Cable
Reservoir Tray
Dust Cover & Tools Kit
Stabilizer 3 KVA
INCLUDE
*.
Instalasi, Training, Guarantee min 1year, Manual operating book, Factory certificate
4 pH meter*)
Specification :
2
*.
pH Range
-2 to 19.999 or wider
*.
mV Range
-1999 to +1999 or wider
*.
Range temperature
*.
*.
pH Relative accuracy
± 0.002 or less
mV Relative Accuracy
± 0.1 or less
*.
Calibration
Max three pH points
*.
Included of
*.
2 set
2 set
190 nm to 700 nm or wider
ANALYTICAL COLUMN C18, 25CM, 5 um
*.
*.
260 x 415 x 500 mm (w x h x d), 27 kg
Sample vial, 1.5 ml (100pc/ pk)
MICRO SYRINGE 100 uL
*.
RSD 0,3 % max
Auto rinsing kit for
*.
*.
*.
Max 1 %
UV-VIS DETECTOR
*.
0.1µL to 100µL
Injection volume accuracy
DEGASSER
*.
Total-volume sample injection, variable injection
volume
4 set
Acessories
d
d
-30.0 C to +130.0 C or wider
2 unit buffer pH 4.01 250 ml : 2 unit buffer pH 7.00
250 ml; 2 unit buffer pH 9.21 250 ml; 2 unit InLab®
Expert Pro "3-In-1" pH electrode
Installation, training min a year, original operation
manual
UNIT
5 Spektrofotometer portable untuk Laboratorium Keliling *)
Specification :
*.
Wavelength range
340-900 nm or wider
*.
Wavelength reproducibility
Wavelength rosolution
0.1 nm or less
1 nm or less
Wavelength Calibration
Internal, automatic at power on, visual feedback
*.
*.
*.
*.
*.
Wavelength accuracy
Spectral Bandwidth
Operating Temperature
Dimensions
216 x 132 x 330 mm
*.
Included of
*.
Acsesories
*.
Pulse swept power
*.
*.
*.
*.
Instalation, Training, Guarantee min. a year, Original
operation Manual
*.
Volume
Touch control keypad
2l
Bath Size
153 x 138 x 100 (l x w x d) mm
Footprint
Net Weight
Include
60W (at 43 kHz ± 2 kHz)
180 x 219 x 179 (l x w x h) mm
2, 4 appr kg
Original operation manual, guarantee min a year
7 Water Deionisation App / Aqua Demineralisator/ Water Puripication *)
Specification : With UV-Lamp
*.
Conductivity
0.1 uS/cm or less
*.
Particles
< 1/ml
*.
*.
*.
*.
TOC
Bacteria
Flow rate
< 6 ppb
< 1 CFU/ml
0,6 l/min
The System includes a pressure pump for supplying equipment connected downstream :
Type I Ultrapure water for Molecular biology, microbiology, PCR, inorganic and organic
trace analysis, HPLC, ICP, TOC-analysis
*.
*.
*.
Type II High purity water for Rinsing laboratory glassware, Preparing and diluting buffers,
reagents, tissue culture media and dyes, Sample preparation for analytical methods such as
flame AAS
Performance
Operating pressure
UNIT
Included : 1-in.square matched glass sample cells;
Cell adapters for 1-inch round/AccuVac cells, 1x1
cells and multi-path 1-inch/1-cm cells; Universal
power supply, 100 to 240V, 47/63Hz, with plug
adapters for EU,GB,US,China; Protective cover for
storing adapters; Dust Cover
Specification
Pengoperasian Penggunaan
1
Line : 100 to 240 V, 47/63 Hz; automatic changeover,
Battery : Lithium-Ion 11V/4400mAh
6 Ultrasonic cleaner *)
*.
UNIT
o
*.
Power Requirements
4
5 nm or less
Source lamp
*.
UNIT
± 1.5 nm
10 to 40 C
Tungsten
*.
1
Min 3 L/ h (output per 8 hours : 24 L) at 15 °C
1
6 bar
Included of UV/UF direct system, 3 Pre-treatment with integrated reverse osmosis ultra pure
water, Sterilizing filter, Ultra filter, UV oxidation 185/254 nm,Tank 6 litres with pressure,
Recirculation pump Instalation, Training, Guarantee min. a year, Original operation Manual
8 Water Distillation App. *)
Specification :
*.
Destilation Capacity
Min 8 L/h
*.
Heating material
Stainless steel or equivalent
*.
Energy saving through distillation of
the heated cooling water
Conform
*.
*.
*.
*.
*.
Conductivity
16 Liter (built in)
Accessories
Pre-filter with filling; *Dechlorite Filter (Cat. #
2904); *Phospate Cartridge (Cat. # 2906); *PreFilter (Cat. # 2912)
Included of
Capacity
Min 45 Liters
*.
Stainless Steel Casing
l x w h : 818 x 516 x 296 mm
*.
*.
*.
*.
*.
*.
Casing, container and cover made of
Interior Heating Concept
Temperature range and control
Electrical supply
220-230 V;50/60 Hz
1 minutes to 99,59 hours or better
Flat Stainless steel cover with opening concentric ring
sets (8 openings); Guarantee min. a year, Original
operation Manual
10 Digestion Apparatus Kjeldahl *)
Specification :
*.
Temperature
*.
Temperature range and control
*.
*.
*.
*.
*.
*.
o
Heating places
Max 600 C
6
Dimensions
600x250x300 (wxdxh) mm
Size of flasks
Electrical supply
Accessories
Included of
UNIT
l x w h : 590 x 350 x220 mm
10 - 95⁰C or wider
LED digital display with PID microprocessor
Included of
1
Stainless steel or equivalent
Control panel
Timer
UNIT
Instalation, Guarantee min. a year, Original operation
Manual
*.
*.
1
220 V, 50/60 Hz, 1 phase, 3.0 KW
9 Waterbath *)
Specification :
UNIT
2.3 µs/cm or less
Storage tank contents
Electrical
1
LED digital display with PID microprocessor
250 mL
220-230 V;50/60 Hz
Water jet pump WSP, 10-0088 1 unit; lsoversinic
tubing 1,2/17 lSO, 10-009L 1 unit
Guarantee min. a year, Original operation Manual
PENGADAAN ALAT LABORATORIUM BPOM DI PALANGKARAYA
TAHUN ANGGARAN 2012
NAMA ALAT DAN SPESIFIKASI
NO
1 Analytical balance*)
Specification :
*.
Capacity
Max 250 g
*.
Reproducibility
0,2 mg
*.
Min piece wight at piece count
*.
*.
*.
*.
*.
Readabilty
Linearity
Overall dimensions (include draft shield)
Weighing plate
Includes
2 Fumehood *)
Instalation, Training, Guarantee min. a year, Original
operation Manual, Dust Cover.
*.
Face velocity 0,5 m.s
*.
*.
*.
*.
Light intensity
Construction Main Body
Power Supply
Accesories
Included of
Min 1720 X 680 X 1200 mm
580 mm, vertical rising sash
Max : 1795 cmh, Min : 900 cmh
780 lux
Phenolic resin Laminate or equivalent
220 - 240 V, 50-60 hz
Base Storage Cabinet with Standard Fittings for 6ft/;
Exhaust Fan/Blower Polypropylene
(diameter :
250mm)
Instalation, Guarantee min. a year, Original operation
Manual
3 HPLC with UV-Vis detector and autosampler*)
Specification :
*. PUMPS
*.
Solvent delivery method
Parallel-type double plunger
*.
Flow rate accuracy
No more than 1% or 0.5uL/min whichever is greater
(0.01 to 2m L/min).
No more than 0.06% or 0.02 min SD. whichever is
greater
*.
*.
Flow-rate setting range
Flow rate precission
Plunger capacity
Maximum discharge pressure
Typical Pulsation
UNIT
80 mm
± 270 mm ( 10.6")
*.
2
WDH : 210 x 330 x 335 mm
Exhaust outlet
Max sash opening
UNIT
0,6 mg/piece
*.
*.
1
Max 0.2 mg
Min 1800 x 875 x 1500 mm
Internal dims
SATUAN
UNIT
± 0.1 mg
Specification :
*. External dims ( LxDxH in mm )
*.
JML
2
2 set
0.0001mL/min. to 10.0000mL/min.
10 uL.
40MPa.
0.03 MPa (for water at 1.0 mL/min and 7 MPa).
Constant-pressure solvent delivery Supported.
*.
*.
Plunger rinsing mechanism
*.
Safety measures
OVEN COLUMN
*.
Temperature range
Manual rinsing or automatic rinsing using optional
product.
Liquid-leakage sensor, high pressure/low-pressure
limits.
4°C to 85°C or wider
2 set
*.
AUTO SAMPLER
*.
Injection method
*.
Injection-volume setting range
*.
Injection volume precision
*.
*.
*.
*.
*.
*.
*.
*.
Number of processed samples
175 (1ml vials)
Dimensions wight
Power requirements
Lamp
Wavelength range
WORKSTATION :
*.
Hardware
*. Software
ACCESSORIS
*.
Maintenance Kit
*.
Mixer
*.
*.
*.
*.
100 VAC, 100 VA, 50/60 Hz
D2 lamp
Intel Core i3-2100 Processor, 2GB 1x2G
CMel/CTino D3-1333, Windows 7 Professional 32bit, HDD 500GB 7200RPMSATA, Intel Graphics
Media Accelerator (GMA) X4500HD, HP USB
Optical Mouse , USB Standard Value Keyboard, HP
15-in-1 Media Card Reader, HP SuperMulti (JB) 1st
ALL, Monitor HP LCD 18,50", Laser Jet HP 1102,
speed up 17 ppm, 1200 dpi effective output, USB 2
port
Data analysis & reporting for food and drug analysis
2 set
2 set
2 set
4 set
4 set
4 set
2 set
2 set
2 set
2 set
2 set
ANALYTICAL COLUMN C8, 15CM, 5 um
HUB W/ Cable
Reservoir Tray
Dust Cover & Tools Kit
Stabilizer 3 KVA
INCLUDE
*.
Instalasi, Training, Guarantee min 1year, Manual operating book, Factory certificate
4 pH meter*)
Specification :
2
*.
pH Range
-2 to 19.999 or wider
*.
mV Range
-1999 to +1999 or wider
*.
Range temperature
*.
*.
pH Relative accuracy
± 0.002 or less
mV Relative Accuracy
± 0.1 or less
*.
Calibration
Max three pH points
*.
Included of
*.
2 set
2 set
190 nm to 700 nm or wider
ANALYTICAL COLUMN C18, 25CM, 5 um
*.
*.
260 x 415 x 500 mm (w x h x d), 27 kg
Sample vial, 1.5 ml (100pc/ pk)
MICRO SYRINGE 100 uL
*.
RSD 0,3 % max
Auto rinsing kit for
*.
*.
*.
Max 1 %
UV-VIS DETECTOR
*.
0.1µL to 100µL
Injection volume accuracy
DEGASSER
*.
Total-volume sample injection, variable injection
volume
4 set
Acessories
d
d
-30.0 C to +130.0 C or wider
2 unit buffer pH 4.01 250 ml : 2 unit buffer pH 7.00
250 ml; 2 unit buffer pH 9.21 250 ml; 2 unit InLab®
Expert Pro "3-In-1" pH electrode
Installation, training min a year, original operation
manual
UNIT
5 Spektrofotometer portable untuk Laboratorium Keliling *)
Specification :
*.
Wavelength range
340-900 nm or wider
*.
Wavelength reproducibility
Wavelength rosolution
0.1 nm or less
1 nm or less
Wavelength Calibration
Internal, automatic at power on, visual feedback
*.
*.
*.
*.
*.
Wavelength accuracy
Spectral Bandwidth
Operating Temperature
Dimensions
216 x 132 x 330 mm
*.
Included of
*.
Acsesories
*.
Pulse swept power
*.
*.
*.
*.
Instalation, Training, Guarantee min. a year, Original
operation Manual
*.
Volume
Touch control keypad
2l
Bath Size
153 x 138 x 100 (l x w x d) mm
Footprint
Net Weight
Include
60W (at 43 kHz ± 2 kHz)
180 x 219 x 179 (l x w x h) mm
2, 4 appr kg
Original operation manual, guarantee min a year
7 Water Deionisation App / Aqua Demineralisator/ Water Puripication *)
Specification : With UV-Lamp
*.
Conductivity
0.1 uS/cm or less
*.
Particles
< 1/ml
*.
*.
*.
*.
TOC
Bacteria
Flow rate
< 6 ppb
< 1 CFU/ml
0,6 l/min
The System includes a pressure pump for supplying equipment connected downstream :
Type I Ultrapure water for Molecular biology, microbiology, PCR, inorganic and organic
trace analysis, HPLC, ICP, TOC-analysis
*.
*.
*.
Type II High purity water for Rinsing laboratory glassware, Preparing and diluting buffers,
reagents, tissue culture media and dyes, Sample preparation for analytical methods such as
flame AAS
Performance
Operating pressure
UNIT
Included : 1-in.square matched glass sample cells;
Cell adapters for 1-inch round/AccuVac cells, 1x1
cells and multi-path 1-inch/1-cm cells; Universal
power supply, 100 to 240V, 47/63Hz, with plug
adapters for EU,GB,US,China; Protective cover for
storing adapters; Dust Cover
Specification
Pengoperasian Penggunaan
1
Line : 100 to 240 V, 47/63 Hz; automatic changeover,
Battery : Lithium-Ion 11V/4400mAh
6 Ultrasonic cleaner *)
*.
UNIT
o
*.
Power Requirements
4
5 nm or less
Source lamp
*.
UNIT
± 1.5 nm
10 to 40 C
Tungsten
*.
1
Min 3 L/ h (output per 8 hours : 24 L) at 15 °C
1
6 bar
Included of UV/UF direct system, 3 Pre-treatment with integrated reverse osmosis ultra pure
water, Sterilizing filter, Ultra filter, UV oxidation 185/254 nm,Tank 6 litres with pressure,
Recirculation pump Instalation, Training, Guarantee min. a year, Original operation Manual
8 Water Distillation App. *)
Specification :
*.
Destilation Capacity
Min 8 L/h
*.
Heating material
Stainless steel or equivalent
*.
Energy saving through distillation of
the heated cooling water
Conform
*.
*.
*.
*.
*.
Conductivity
16 Liter (built in)
Accessories
Pre-filter with filling; *Dechlorite Filter (Cat. #
2904); *Phospate Cartridge (Cat. # 2906); *PreFilter (Cat. # 2912)
Included of
Capacity
Min 45 Liters
*.
Stainless Steel Casing
l x w h : 818 x 516 x 296 mm
*.
*.
*.
*.
*.
*.
Casing, container and cover made of
Interior Heating Concept
Temperature range and control
Electrical supply
220-230 V;50/60 Hz
1 minutes to 99,59 hours or better
Flat Stainless steel cover with opening concentric ring
sets (8 openings); Guarantee min. a year, Original
operation Manual
10 Digestion Apparatus Kjeldahl *)
Specification :
*.
Temperature
*.
Temperature range and control
*.
*.
*.
*.
*.
*.
o
Heating places
Max 600 C
6
Dimensions
600x250x300 (wxdxh) mm
Size of flasks
Electrical supply
Accessories
Included of
UNIT
l x w h : 590 x 350 x220 mm
10 - 95⁰C or wider
LED digital display with PID microprocessor
Included of
1
Stainless steel or equivalent
Control panel
Timer
UNIT
Instalation, Guarantee min. a year, Original operation
Manual
*.
*.
1
220 V, 50/60 Hz, 1 phase, 3.0 KW
9 Waterbath *)
Specification :
UNIT
2.3 µs/cm or less
Storage tank contents
Electrical
1
LED digital display with PID microprocessor
250 mL
220-230 V;50/60 Hz
Water jet pump WSP, 10-0088 1 unit; lsoversinic
tubing 1,2/17 lSO, 10-009L 1 unit
Guarantee min. a year, Original operation Manual