VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
7 VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
Y OSHIKI N ISHIYAMA Eli Lilly Japan K.K.
7.1 INTRODUCTION
In this chapter we describe characteristic validation procedures of the Heavy Met- als Limit Test in the Japanese Pharmacopoeia (JP) [1]. Although an equivalent test is commonly listed in both the United States Pharmacopoeia and the Euro- pean Pharmacopoeia , there are differences in the color reagents and conditions of sample preparation of the JP procedure. Heavy metals are defined in the JP as poisonous metallic impurities such as Pb, Bi, Cu, Cd, Sn, and Hg that form colored colloidal precipitates with sodium sulfide TS in a slightly acidic solution of pH 3 to 4. The level is expressed as the equivalent quantity of lead.
In JP monographs, the specification value and testing procedure are described as a comparative limit test for the quantity of heavy metals that exist in drug substances as inorganic impurities. The permissible limit on the “ppm” scale for heavy metals (as Pb) is prescribed as the specification value. From the devel- opment stage to the establishment of the limit test method for a monograph, the validation in place has to be done for both the quantitative method for actual measurement and the comparative limit method with the control solu- tion. However, the quantitative procedure has not been stated in the section on the heavy metals limit test in the JP’s General Tests and other Japanese official
Analytical Method Validation and Instrument Performance Verification , Edited by Chung Chow Chan, Herman Lam, Y. C. Lee, and Xue-Ming Zhang ISBN 0-471-25953-5 Copyright 2004 John Wiley & Sons, Inc.
96 VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
guidelines, even though the validation package for the actual value measurement is usually required as the rationale of limit value described in regulatory sub- mission documents. Therefore, in this chapter we describe the approaches of validation studies that are required to establish and implement practices concern- ing the quantitation and the limit testing methods for heavy metals.
We focus on the requirements for registration in Japan. Thus, we provide information that considers the peculiarities of the requirements for the JP heavy metals test at the method development process. The compendial descriptions and relevant ICH guidelines in relation to this chapter are listed in the reference section. The principles for the validation requirements discussed in this chapter can be applied to heavy metal testing in general.
7.2 SCOPE OF CHAPTER
In the process for validation studies for developing the limit test of heavy metals according to JP requirements for the specifications of new drug substances, there are usually two significant individual procedures: (1) validation for the quan- titation of heavy metals, and (2) validation for the limit test of heavy metals. Validation of the quantitation for heavy metals test assesses the applicability of a quantitation method where the measured values serve as the basis for the limit test used for the specification of heavy metals in the drug substance. This method determines the concentration of lead by a UV–Vis spectrophotometer using a calibration curve. These data should be described in regulatory submis- sion documents as a piece of essential information for a new drug application. We follow with a discussion on characteristics and procedures of validation for developing the limit test specified in “Specifications and Testing Methods” of the drug substance.
7.3 VALIDATION PRACTICES
In this chapter, validation studies for developing the quantitative and JP limit test methods for heavy metals described in “Specifications and Testing Methods” of the drug substance section are described. Procedures to evaluate validation characteristics concerning the validation items of each test method are described. Table 7.1 summarizes the validation characteristics required by JP for testing of impurities.
7.3.1 Validation Procedures of Quantitation of Heavy Metals
The quantitation of heavy metals in a drug substance is a method to determine the quantity of lead by determining the intensity of coloring with sodium sulfide based on the absorbance at 400 nm by spectrophotometry. It is a prerequisite to set up the limit test according to the JP requirements. With regard to validation characteristics of the quantitation method, specificity is not required because of
97 Table 7.1. JP Validation Characteristics Requirements a
VALIDATION PRACTICES
Testing for Impurities Assay (dissolution;
Characteristic
Limit content/potency) Accuracy
Identification
– + b Intermediate precision
+ + Detection limit
Specificity c +
– d + – Quantitation limit
– – Linearity
– + Range
– + a −Signifies that this characteristic is not normally evaluated; + signifies that this characteristics is normally evaluated.
b Where reproducibility has been established, intermediate precision is not needed. c Lack of specificity of one analytical procedure could be compensated by supporting analyti-
cal procedure(s). d May be needed in some cases.
the definition of heavy metals as described in “Heavy Metals Limit Test” section of JP general testing method 22. Consequently, the following four validation characteristics should be evaluated: (1) accuracy, (2) precision (repeatability), (3) range and linearity, and (4) quantitation limit. Accuracy and precision are evaluations on the appropriateness of procedures adopted from the test solution preparation of methods 1 to 4 in JP. The sample content must be evaluated so that it is within the linearity, range, and quantitation limit.
Accuracy. In the quantitative method that is used to measure the heavy metal quantity in the drug substance, the accuracy is usually represented by the recovery rate obtained from a spiked recovery test where lead is added to the samples. Since the heavy metals limit test specified in monograph specifications is a test where the intensity of coloring of the samples with sodium sulfide is compared with that of the control solution, it is necessary to confirm that heavy metal components can be detected fully in the process of test solution preparation. The “Heavy Metals Limit Test” in JP specifies four preparation methods for the test solutions. An appropriate method will be selected and used for further testing. The test method that gives the best recovery rate is to be adopted. The procedure is as follows:
Procedure
1. Add a volume of standard lead solution containing each of the following three levels of lead to samples of the drug substance: the content equal to the expected specification limit and those higher and lower than the
98 VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
median level. At least three spiked samples should be prepared for each lead content.
2. Treat the three-level lead-spiked drug substance samples according to meth- ods 1 to 4 to prepare the test solutions and the control solution. Separately, designate a solution prepared according to the same preparation method as the control solution. Using this control solution, determine the absorbance of the test solutions and calculate the recovery rate at each amount added.
A typical example of accuracy (recovery rate) is shown in Table 7.2. The samples were prepared by adding a standard lead solution corresponding to 2, 5,
10, and 15 ppm to 2.0 g of a compound. Results of the recovery were determined from the absorbance at 400 nm.
Precision (Repeatability). To evaluate the repeatability as specified in the quan- titative method of heavy metals in the drug substance, the drug substance samples are treated according to the test solutions and the control solutions preparation method selected from methods 1 to 4 of “Heavy Metals Limit Test” in JP. Take five or six aliquot samples collected from a single lot of homogeneous drug sub- stance and determine the quantity of heavy metal in each sample aliquot using the prepared test and control solutions. The data obtained are statistically analyzed.
Procedure
1. Take five or six aliquot samples from a single lot of homogeneous drug substance. The sample amount is equal to that specified in the limit test method in the “Specifications and Testing Methods.”
2. Prepare the test and control solutions from the aforementioned sample according to the method of test solutions selected and the control solutions preparation procedure specified in the limit test method in the “Specifica- tions and Testing Methods.” Use these solutions to determine the quantity of heavy metal by the same calibration curve from the accuracy experiment. This procedure is repeated for each test solution.
3. Perform statistical analysis to obtain the standard deviation and relative standard deviation.
Table 7.2. Recovery of Heavy Metals (%)
Amount of Lead Added (ppm) Replicate
Average 94.3 94.4 91.2 92.2
99 Table 7.3. Testing for Repeatability
VALIDATION PRACTICES
Value Observed
A typical example of precision (repeatability) is shown in Table 7.3. The sam- ples were prepared from 2.0 g of a compound. The data obtained were determined from the absorbance at 400 nm.
Range and Linearity. The range for which the linearity of the absorbance ver- sus lead concentration calibration curve applied for the heavy metal quantitative method has to be confirmed. A wide range from ca. 1 ppm to two times the specification limit value (e.g., 15 or 20 ppm concentration relative to the sam- ple weight) should be used for the evaluation range of the calibration curve. In general, it is recommended to use at least five concentration levels at regular intervals by using lead standard solution within the range studied. Actual data obtained are analyzed statistically by obtaining a regression line using the least squares method to evaluate linearity within the range specified. The linearity of the calibration curve is evaluated by the correlation coefficient, y-intercept, and slope of the regression line. A plot of the regression line should be included.
Procedure
1. Place an appropriate volume of standard lead solution which corresponds to each of five concentration levels in a Nessler tube, and add water to make up to 40 mL. Add 2 mL of dilute acetic acid and water to make up to 50 mL, and designate it as the test solution. Three tubes of test solution at each concentration level are prepared for separate runs.
2. The control solution is prepared by pipetting 2 mL of dilute acetic acid in
a Nessler tube of the same material with the test solution and adding water to make up to 50 mL.
3. Add a drop of sodium sulfide to the test and control solutions, mix, and leave for 5 min. Then determine the absorbance of the test solution using the control solution as the control value.
4. Determine the absorbance at each concentration, and obtain the regression curve of lead concentration versus absorbance by the least squares method.
Figure 7.1 illustrates the typical linearity obtained for this test. The samples were prepared by adding standard lead solution corresponding to 2, 5, 10, and
15 ppm to 2.0 g of a compound. The response was linear within the range.
VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
0.02 Slope (AU/ppm): 0.00146
y -intercept (AU): 0.0000 0.01 Correlation coefficient: 0.9997
Absorbance at 400 nm
Concentration of standard lead solution (ppm)
Figure 7.1. Linearity of calibration curve.
Quantitation Limit. When the quantity of heavy metals is determined from the calibration curve, it is recommended to estimate the lowest value of heavy metals concentration as the quantitation limit. The methods to estimate the quantitation limit are described in JP and ICH Q2B Guidelines [2], and an appropriate method should be selected from among these methods. An estimation of the quantitation limit can be obtained from the standard deviation of measured values of the low- concentration test solution. The standard deviation of background noise will be used to estimate a value for the standard signal-to-noise ratio (10 : 1).
Procedure
1. Prepare five to six samples of the test solution with known low-lead concen- tration that is close to the quantitation limit, and determine the absorbance at 400 nm by spectrophotometer.
2. Calculate the standard deviation of the samples above. Use this standard deviation as the standard deviation of background noise.
3. Assuming that the standard deviations of the measured heavy metals values maintain the linearity up to the limit value, the quantitation limit (QL) cor- responding to the signal-to-noise ratio of 10 : 1 (RSD = 10%) is calculated using the following formula:
QL = SD × 10
where SD is the standard deviation of the absorbance of the test solution with known low-lead concentration that is close to the quantitation limit.
Using this procedure, the estimated quantitation limit is obtained from the variation of actual measurement values of low-concentration test solutions instead of background noise. However, the actual value of quantitation limit needs to be verified by experiment.
VALIDATION PRACTICES
7.3.2 Validation Procedures of Limit Test of Heavy Metals
The validation of heavy metals limit test method is required to evaluate the specificity and the detection limit of the limit test of impurities. However, since the JP method is capable of detecting Pb, Bi, Cu, Cd, Sn, and Hg, specificity evaluation for these metals is not required. The detection limit of the method should be evaluated visually.
Detection Limit. The detection limit of heavy metals limit test method is obtained from the test solutions and the control solution. These solutions are prepared using one of methods 1 to 4 of the control solution preparation method. The detection limit is determined by visual inspection of a series of diluted standard lead solutions.
Procedure
1. Adjust the volume of lead standard solution to be the same as that for preparation of the control solution. Prepare at least seven or eight levels of lower-concentration test solutions from the specification limit value of 0.1 to 0.2 ppm (single replicate).
2. Add a drop of sodium sulfide TS to the test solutions, mix, and hold for 5 min. Put the Nessler tube containing the test solution of each lead concentration on a white background. Visually observe it from the upper side. Determine the lowest concentration where darkening can be visually observable and designate it as the detection limit.
7.3.3 Preparation of Test Solutions and Control Solutions
The following are citations from the section on the preparation of test solutions and control solutions in JP 14, “Heavy Metals Limit Test.” Unless otherwise specified, test solutions and control solutions are prepared as directed below:
Method 1. Place an amount of the sample, directed in the monograph, in a Nessler tube. Dissolve in water to make 40 mL. Add 2 mL of dilute acetic acid and water to make 50 mL, and designate it as the test solution. The control solution is prepared by placing the volume of standard lead solution directed in the monograph in a Nessler tube, and adding 2 mL of dilute acetic acid and water to make 50 mL.
Method 2. Place an amount of sample, directed in the monograph, in a quartz or porcelain crucible, cover loosely with a lid, and carbonize by gentle ignition. After cooling, add 2 mL of nitric acid and 5 drops of sulfuric acid, heat cautiously until white fumes are no longer evolved, and incinerate by ignition between 500 and 600 ◦
C. Cool, add 2 mL of hydrochloric acid, evaporate to dryness on a water bath, moisten the residue with 3 drops of hydrochloric acid, add 10 mL of hot water, and warm for 2 min. Then add 1 drop of phenolphthalein TS, add ammonia TS dropwise until the solution develops a pale red color, add 2 mL of dilute acetic acid, filter if necessary, and wash with 10 mL of water. Transfer the filtrate and washings to a Nessler tube, and add water to make 50 mL. Designate it as the test solution.
VALIDATION STUDY OF JP HEAVY METAL LIMIT TEST
The control solution is prepared as follows: Evaporate a mixture of 2 mL of nitric acid, 5 drops of sulfuric acid, and 2 mL of hydrochloric acid on a water bath, further evaporate to dryness on a sand bath, and moisten the residue with
3 drops of hydrochloric acid. Proceed as directed in the test solution, then add the volume of standard lead solution directed in the monograph and water to make 50 mL.
Method 3. Place an amount of the sample, directed in the monograph, in a quartz or porcelain crucible, heat continuously, gently at first, and then increase the heat until incineration is completed. After cooling, add 1 mL of aqua regia, evaporate to dryness on a water bath, moisten the residue with 3 drops of hydrochloric acid, add 10 mL of hot water, and warm for 2 minutes. Add 1 drop of phenolphthalein TS, add ammonia TS dropwise until the solution develops a pale red color, add 2 mL of dilute acetic acid, filter if necessary, wash with 10 mL of water, transfer the filtrate and washing to a Nessler tube, and add water to make 50 mL. Designate it as the test solution.
The control solution is prepared as follows: Evaporate 1 mL of aqua regia to dryness on a water bath. Proceed as directed for the test solution, and add the volume of standard lead solution directed in the monograph and water to make 50 mL.
Method 4. Place an amount of the sample, directed in the monograph, in a plat- inum or porcelain crucible, mix with 10 mL of a solution of magnesium nitrate hexahydrate in ethanol (95) (1 in 10), fire the ethanol to burn, and carbonize by gradual heating. Cool, add 1 mL of sulfuric acid, heat carefully, and incinerate by ignition between 500 and 600 ◦
C. If a carbonized substance remains, moisten with
a small amount of sulfuric acid, and incinerate by ignition. Cool, dissolve the residue in 3 mL of hydrochloric acid, evaporate on a water bath to dryness, wet the residue with 3 drops of hydrochloric acid, add 10 mL of water, and dissolve by warming. Add 1 drop of phenolphthalein TS, add ammonia TS dropwise until
a pale red color develops, then add 2 mL of dilute acetic acid, filter if necessary, wash with 10 mL of water, transfer the filtrate and the washing to a Nessler tube, add water to make 50 mL, and use this solution as the test solution.
The control solution is prepared as follows: Take 10 mL of a solution of magnesium nitrate hexahydrate in ethanol (95) (1 in 10), and fire the ethanol to burn. Cool, add 1 mL of sulfuric acid, heat carefully, and ignite between 500 and 600 ◦
C. Cool and add 3 mL of hydrochloric acid. Proceed as directed in the test solution, then add the volume of standard lead solution directed in the monograph and water to make 50 mL.
7.4 COMMON PROBLEMS AND SOLUTIONS
1. How should the criteria for selections from methods 1 to 4 of the test be established, and the solution preparation specified in the heavy metals limit test in JP be controlled?
REFERENCES
103
ž Method 1 can be used if the drug substance can be dissolved in 40 mL of water. There is no precipitation after adding 2 mL of dilute acetic acid. The pH of the solution can easily be adjusted to 3.0 to 3.5. It results in a favorable recovery rate.
ž Method 2 is commonly used, while method 3 is rarely used for items listed in JP. Method 4 should be used when an unsatisfactory recovery rate is obtained with method 2.
2. The spiked recovery test cannot produce a favorable result. The following precautions should be taken at the time of preparation of the test and control solutions from samples.
ž Carbonize by gradual ignition to avoid burning of the sample or loss by swelling outside the crucible.
ž For incineration with sulfuric acid in methods 2 and 4, be sure to limit the temperature to 500 to 600 ◦
C to avoid volatilization of heavy metals. ž Ensure that the pH of the test solution is in the range 3.0 to 4.0 using dilute
acetic acid during preparation. ž In the case of method 4, use a slightly larger crucible to prevent scattering of the sample during burning with ethanol.
3. After adding sodium sulfide TS, turbidity is observed when the test and control solutions are compared.
ž It is recommended to add sodium sulfide reagent solution to the test and control solutions at the same time and to observe them approximately 5 min later. After that time, sulfur will precipitate causing turbidity.
4. Even after addition of sodium sulfide TS and mixing well, there is only slight darkening, or there is existing coloring in the reagent solution.
ž Sodium sulfide TS contains glycerin to stabilize and inhibit oxidation. The ability to produce sulfide is decreased with storage time. It is recommended that the fresh reagent solution should be used, especially for the quantitation test. Even if the reagent solution is refrigerated, it should be used within a couple of months.