LPSE POM

PANITIA PENGADAAN BARANG DAN JASA
PENGADAAN ALAT LABORATORIUM
BALAI BESAR PENGAWAS OBAT DAN MAKANAN DI BANJARMASIN
ADDENDUM II LAMPIRAN SPESIFIKASI TEKNIS
PENGADAAN ALAT LABORATORIUM
No. 01.02.100.04.12.0700
Tanggal 18 April 2012
Sehubungan dengan perubahan, penambahan dan atau pengurangan dari Dokumen Pengadaan untuk Pekerjaan Pengadaan Alat Laboratorium
No. PR.01.02.100.04.12.0631 tanggal 9 April 2012 terdapat perubahan pada spesifikasi teknis sebagai berikut :
1. No. 5. Aqua Reverse Osmosis System
Built in reservoir volume : 6 L
Point g.
dihilangkan
Terlampir Spesifikasi Teknis Pengadaan Alat Laboratorium setelah addendum 1 dan 2

Lampiran Spesifikasi Teknis
Dok. 01.02.100.04.12.0631
PENGADAAN ALAT LABORATORIUM BALAI BESAR POM di BANJARMASIN
RINCIAN SPESIFIKASI TEKNIS PENGADAAN ALAT TAHUN 2012
No.


SPESIFIKASI

1 HPLC with auto & manual injector, UV-VIS & PDA Detector *)
Specifications :
Pompa
- System Pump
: Binary and Quaternary

SATUAN

JUMLAH

Unit

1

- Flow rate setting range
: 0.001mL/min. to 10 mL/min.
- Flow rate accuray
: ≤ %

- Flow rate precission
: ≤ . %
: minimal 4
- Number of solvent mixed
- Mixing - concentration accuracy
: ≤ . %
Detektor uv-vis
- Light source
: D2 lamp
- Wavelength range
: 190-700 nm or wider
- Wavelength accuracy
: ± 1 nm
- Wavelength precision
: ± 0,1 nm
- Funtion
: Dual Wavelenght
- Linearity
: 2.5 AU
Detektor Photo Array Detector

- Light source
: D2 lamp, Tungsten (W) lamp
- Number of diode elements
: min 512
- Wavelength range
: 190-800 nm or wider
- Wavelength accuracy
: ± 1 nm
- Wavelength precision
: ± 0,1 nm
- Funtion
: contour output, spectrum library
- Linearity
: 2.0 AU
Column Oven
- Temperature setting range
: 4°C to 80°C or wider
- Temperatur control precision
: ≤ , oC
- Coulumn storage

: min 4 columns
Autosampler
- Injection Precision
: < 0,5% or better
- Injection volume
: 0,1-100 uL
- Number of sample vials
: minimal 100 vial@±1-2 ml
- Sample carryover
: min 0.01%
Manual injector
: 1 unit
Can inject from 1 µl to 5 ml with high accuracy and precision
System Controller
Number of connectable units
: min 8 units

Computer
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor

LCD 18,50", Laser colour printer, USB 2 port.
Data Analysis & Reporting for food and drug analysis.
Include :
a.
b.
c.
d.
e.
f.
g.
h.
i.
j.
k.
l.
m.
n.
o.
p.
q.


Maintenance Kit for Pump
Auto rinsing kit for Pump
Mixer SUS
Sample vial, 1 - 2ml (100pc/ pk) min 2 pkg
Micro syringe 100 uL 1 pcs
C18 Coulumn 4.6x250 mm, 5 um min 3 pcs
C 8 Coulumn 4.6x250 mm, 5 um min 2 pcs
Check valve inlet min 2 pcs
HUB W/ Cable
Tools Kit
UPS 5 KVA
Reservoir Tray
Instalasi
Training
Original manual book
Factory cerificate
Guarantee min 1 year

2 HPLC with auto & manual injector, UV-VIS Detector with manual loop 500µl *)

Specifications :
- System Pump
: Binary and Quaternary
- Flow rate setting range
: 0.001mL/min. to 10 mL/min.
- Flow rate accuray
: ≤ %
- Flow rate precission
: ≤ . %
: minimal 4
- Number of solvent mixed
- Mixing - concentration accuracy
: ≤ . %
Detektor uv-vis

Unit

1

- Light source

: D2 lamp
- Wavelength range
: 190-700 nm or wider
- Wavelength accuracy
: ± 1 nm
- Wavelength precision
: ± 0,1 nm
- Funtion
: Dual Wavelenght
- Linearity
: 2.5 AU
Column Oven
- Temperature setting range
: 4°C to 80°C or wider
- Temperatur control precision
: ≤ , oC
- Coulumn storage
: min 4 columns
Autosampler
- Injection Precision

: < 0,5% or better
- Injection volume
: 0,1-100 uL
- Number of sample vials
: minimal 100 vial@±1-2 ml
- Sample carryover
: min 0.01%
Manual injector
Can inject from 1 µl to 5 ml with high accuracy and precision
System Controller
Number of connectable units
: min 8 units

Computer
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor
LCD 18,50", Laser colour printer, USB 2 port
Data Analysis & Reporting for food and drug
Include :
a.

b.
c.
d.
e.
f.

Maintenance Kit for Pump
Auto rinsing kit for Pump
Mixer SUS
Sample vial, 1 - 2ml (100pc/ pk) min 2 pkg
Micro syringe 100 uL 1 pcs
C18 Coulumn 4.6x250 mm, 5 um min 2 pcs

g.
h.
i.
j.
k.
l
m.

n.
o.
p.
q.
r.
s.

C 8 Coulumn 4.6x250 mm, 5 um min 2 pcs
C18 Coulumn 4.6x150 mm, 5 um min 1 pcs
Check valve inlet min 2 pcs
HUB W/ Cable
Tools Kit
UPS 5 KVA
Reservoir Tray
Sample loop 500µl for manual ijnector
Instalasi
Training
Original manual book
Factory cerificate
Guarantee min 1 year

3 HPLC with autosampler, UV-VIS Detector *)
Specifications :
Pompa
- System Pump
- Flow rate setting range
- Flow rate accuray
- Flow rate precission
- Gradient for easy method development
- Number of solvent mixed
- Mixing - concentration accuracy
Detektor UV-Vis
- Light source
- Wavelength range
- Wavelength accuracy
- Wavelength precision
- Flow cell design

- Funtion

Unit

:
:
:
:
:
:
:
:
:
:
:
:

Quaternary
0.000 mL/min. to 10.000 mL/min. (0.050 to 5.000 mL/ min typical)
≤ %
≤ . %
min 10 gradient curves function
minimal 4
≤ . %

D2 lamp
190-700 nm or wider
± 1 nm
± 0,1 nm
maximizes energy throughtput, eliminates refractive index effect,
ensuring the highest sensitivity while maintaining optimal spectral
performance
: Dual Wavelenght

2

- Linearity
- Lamp optimization

Column Oven
- Temperature setting range
- Coulumn storage
Autosampler
- Injection volume
- Number of sample vials
- Sample carryover
- Injection Precision

: 2.5 AU
: Designed to automatically maximize signal to noise in both visible
and UV spectrum with the use of the single D2 lamp, extends the
useful life of the lamp for consisten result overtime
: 20°C to 65°C or wider
: min 2 columns
: 0,1-100 uL
: minimal 100 vial@±1-2 ml
: min 0.1%
: < 0,5% or better

Computer
Min Intel Core i3-2100 Processor, 4GB DDR3, Windows 7 Professional 32-bit, HDD 500GB, DVD RW, Intel
Graphics Media Accelerator (GMA) X4500HD,intel G-41 chipset, USB Optical Mouse , USB Keyboard, , Monitor
LCD 18,50", Laser colour printer, USB 2 port
Data Analysis & Reporting for food and drug analysis.
Include :
a.
b.
c.
d.
e.
f
g.
h.
i.
j.
k.

Tool kit
C18 Coulumn 4.6x250 mm, 5 um min 2 pcs
C18 Guard Coulumn 4.6x20 mm, 5 um min 2 pcs
Sample/standard filtration, Ø 13 mm 0.45 µm
Sample vial, 1 - 2ml (100pc/ pk) min 1 pkg
UPS 2 KVA
Instalasi
Training
Original manual book and software
Factory cerificate
Guarantee min 1 year

4 Aquadestillata (Glass-Destilation) *)
Spesification

unit

1

a.
b.
c.
d.
e.
f.
g.
h.
i.
j.
k.
l.
m.
n
o.
p.
q.

Included of
-

Capacity
Cooling water requirement
Conductivity
Evaporator material
Condensor material
Heating element material
Low water cut off
Energy saving
Distillate withdrawl
Cooling water outlet
Impurity detector
Housing
Cooling water pressure
Dimension (WxDxH)
Power input
Electrical consumption
Another features

:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:
:

min. 8 L/ hour
± 144 L/ hour
± 2.2 µs/cm at 20°C
borosilicate glass
borosilicate glass
Quartz sheathing
in condensor, auto power cut off
through distillation of heated cooling water
Right hand side
Ø 12.7 mm
Automatic
Galvanized steel, powder coated with epoxy resin
> 3-7 bar
max 650 x 365 x 390
400 V/ 50-60 Hz
max. 6.0 kW
- Electronic impurity detector by automatic water change
- Water saving through automatic water cut-off
- Sterillisation of the condensers by steam
- CO2 degassing through outlet in the condenser
- Water supply and cooling water outlet with hose
connection 1/2 inch.

Wall or bench mounting
Instalation
Guarantee min 1 year
Original operational manual book

5 Aqua reverse Osmosis System *)
Spesification
a.
Nominal permeate flow
b.
Product water volume
b.
Inlet feed water

unit
: min 16 L/h
: 160 to 320 L/day
: tap water

1

c.
d.

Ionic content rejection
Organics, particulate
microorganism rejection
e.
Dimension (HxWxD)
f.
Operating weight
g.
Electrical feed voltage
h.
Tap (feed) water connection
i.
Tap (feed) water pressure
j.
Water quality
Accessories/catridge used
RO membrane catridge (2 pcs)
Sanitization chlorine tablet (2 pack)
Vent filter for PE reservoir (2 pcs)
Pretreatment kit (2 pcs)
100 L PE reservoir
(1 pc)
1200 L tank
(1 pc)
Water pump (1 pc)
Included of
Installation
Guarantee min. 1 year
Original operational manual book
6 Thermocouple datalogger with printer
Spesification
a.
Temperature Range & accuracy
b.
Resolution
c.
Class
d.
Display
e.
Environment operation
f.
Measurement channels
g.
Display of temperature
h.
Source of power

: > 96 %
: > 99 %
:
:
:
:
:
:

max. 45.7 x 25.5 x 31.5 cm
max 15 kg
220 - 240 V
1/2" Gaz M
0.5 - 6 bar
type 3

unit
:
:
:
:
:
:
:
:

min. -200-1370o C ± 0.3 %rdg + 0.7 o C
max. 0.1 o C
K
Digital, triple display T1, T2, T1-T2
o
- 20-50 C ; RH < 90%
min.2
in degrees Celcius or Fahrenheit
AA battery

2

i.

j.

Include
:
thermocouple probe (2 unit) which compatible with apparatus and measured temperature
calibration features
printing paper rol
Include
Guarantee min. 1 year
Original operation manual

7 Biosafety Cabinet *)
Specifications
a.
Type
b.
External dimension
c.
Main control
d.
Air flow velocity
e.
Display
f.
Main filter
g.
Cabinet construction
h.
Sound emision
i.
Power
Incuded of :
- UV Lamp
- Support Stand
- Electric outlet
- Instalation
- Training
- Guarantee min. 1 year
- 0riginal operation manual
8 Incubator *)
Specification :
a.
Capacity
b.
Temperature range
c.
Resolution

:
:
:
:
:
:
:
:
:

Unit

1

unit

1

class II
max 1420 x 815 x 1540 mm
Digital microprocessor control system
Inflow 0.45 m/s
digital display
ULPA filter effisiency min : 99.999%
electrogalvanized steel with antomicrobial coating
< 53.5 dBA (EN 12469) &