Validasi metoda november 2016 new
Recent development on Validation Method of
Chemical Analysis for (Phyto)Pharmaceutical Preparations
Gunawan Indrayanto
Plant Biotechnology Research Group,
Faculty of Pharmacy, Airlangga University
Surabaya 60286, Indonesi
a
gunawanindrayanto@yahoo.com
This presentation is for Education or Training only;
this presentation is not for commercial purposes
This presentation can not be copied without any permission from the author
Some slides were cited/reproduced from free sources website
(2)
Why chemical/Pharmaceutical analysis
is very important?
Chemical and Pharmaceutical Analysis:
•
Pharmaceutical/ Chemical/Food/Cosmetic Industries
•
Diagnostic Laboratories
•
Forensic Laboratories
•
Government Laboratories (BPOM, FDA etc)
•
Government Laboratories (BPOM, FDA etc)
•
Research
How can we have a reliable results of the
chemical analysis?
If the results is NOT valid, all conclusions
will be NOT correct.
(3)
Areas
of Chemical Analysis
•
Quantification:
–
How much
of substance X, Y, Z etc is in the
sample?
•
Detection:
–
Does the sample
contain
substance X, Y, Z etc?
–
Does the sample
contain
substance X, Y, Z etc?
•
Identification:
–
What is the
identity
of the substance(S) in the
sample?
•
Separation:
–
How can the species of interest
be separated
from
the sample matrix for better isolation,
quantification and identification?
(4)
“Valid”
Methods
Calibrated
Software
?
Reference
Validation
method
compounds
Target compound
must be
stable
in the
Results
of
Analysis
Conclusion(s)
System
Suitability check
Routine Analytical Method
IQ,
OQ, PQ
are valid
Calibrated
Instruments
Qualified
Person
must be
stable
in the
selected solvent
(5)
E. Rozet, Ph. Hubert, Presentation University de Liege, Tiblisi, August 2011
(6)
(7)
(8)
(9)
Reliable
and
Validated
In order to determine
the “Quality”
of any (Herbal) Drugs,
all chemicals
(API, metabolites, impurities, degradation
products, heavy metals, pesticides, preservatives, sweetener,
toxin) must be evaluated (
qualitatively and quantitatively
).
analysis’s
methods are
needed !
(10)
Validation method
for specific compound(s):
drugs, metabolites,
drugs, metabolites,
markers, metals, pesticides,
preservatives
(11)
(12)
Farmakope Indonesia V
(13)
(14)
(15)
(16)
Acceptance criteria of accuracy and
precision are depended
(For compound/marker approach)
depended on Pre-determined Specification Range Or Analytical Target Profile (ATP)
(17)
Massart’s Golden rules of Validation method
(Cited by Gonzales & Herrador,
(18)
Data Elements Required for Analytical Method Validation
Type of Analytical Procedure
Assay Impurity testing Quantitative Limit tests
Performance charcteristics
Identification
Parameter : USP ICH USP ICH USP ICH USP ICH USP ICH
Yes * *
(For Drugs/Cosmetic/Food analysis)
Accuracy Yes Yes Yes Yes * No * n.a No No
Precision Yes n.a Yes n.a No n.a Yes n.a No n.a
Repeatability n.a Yes n.a Yes n.a No n.a n.a n.a No Interm. Prec. n.a Yes n.a Yes n.a No n.a n.a n.a No
Specificity Yes Yes Yes Yes Yes Yes * n.a Yes Yes
DL No No No No Yes No * n.a No No
QL No No Yes Yes No No * n.a No No
Linearity Yes Yes Yes Yes No No * n.a No No
Range Yes Yes Yes Yes * No * n.a No No
(19)
Step taken for validation Process
Analytical goal
Method selection (Published / new)
Optimization
Stability testing (SST and pre-validation)
Stability testing (SST and pre-validation)
Testing validation parameters
Validation report
Transfer of Method (R&D to QC)
(20)
(21)
(22)
For analysis using (Chromatography)
SST
must be determined
( for standard/ QC samples):
According to USP /39/40 FI V the parameters are:
•Rs (usually > 1.75)
•N
•Tailing Factor ( nicer if ca, 1)
Requirements for doing verification of an official method
•Tailing Factor ( nicer if ca, 1)
•S/N ratio (>5)
•Replicate Injections: see each of monographs or
% RSD (maximal)=
(23)
If B= 5, n = 6, max RSD = 2.17 % B = 4, n = 6, max RSD = 1.39 %
(24)
(25)
(26)
Definition of Markers (2008)
:Not always = maybe “yes”
maybe “no”
(27)
(28)
Concentrations of
Selected (important) markers could
determine the activities
(29)
Must be determined
(30)
Stability Studies
Stability Studies
(31)
Pre-Validation: Stability testing
•
Scope:
Usage of autosampler for overnight runs with samples
in solution for hours in laboratory environment raises
concern about the stability of samples.
e.g., degradation by hydrolysis, photolysis, adhesion to
glassware, etc.,
•
Recommendations:
•
Recommendations:
Generate data to support standard and sample solution
stability under normal lab conditions for duration of
test procedure, e.g., 24 hours, 3-5 hours for TLC Profile
Acceptance Criteria
Maximal peak area difference is ± 2 %, compare to the
fresh solution
(32)
(33)
(34)
Pre-Validation: Stability data of reference compound and sample should be reported Contoh: Analisa dengan HPLC
(35)
(36)
Reference standards:
originality (USP, BP, SIGMA, MERCK, etc);(37)
Instrumental analysis (chromatography & spectroscopy
need specific authentic r
standards/compounds
except, if we use Q-NMR (need only general IS)(38)
(39)
Purity of the standards
-Using mass Balance Method
(40)
Mass balance for primary standard;
(It need (a) calibrated micro/analytical balances)
(41)
Smallest minimum weight
(42)
Smallest
net weight (w)
according USP: 38 /2015; 39/2016
Sr =
Sr = repeatability ( n = 10) = not more than 0. 10%
W =
Sr = repeatability ( n = 10) = not more than 0. 10%
d = scale interval
For analytical balance d = 0.1 or 0.01mg ; micro-balance 0.001 mg,
w will be = 82 mg , 8.2 mg
,
0.82 mg
(if SD = 0 or , < 0.41 d)
Or w = 2000 SD
USP 35/FI V : Tolerance not exceed 0. 1 % of the amount weight;
so smallest weight must be 100.0 (4 digits)and 10.00 mg (5 digits), or
1.000 mg (6 digits).
(43)
(44)
(45)
Selectivity/Specificity
(46)
Specificity/Selectivity
USP, ICH, FDA: Specificity
AOAC, IUPAC : Selectivity
Specificity is
the ability to determine
the analyte unambiguously
in the presence of other components, this includes degradation
products, metabolites, matrix component etc.
products, metabolites, matrix component etc.
Selectivity means that the method provides responses for a
number of chemical entities that may
distinguished
from each
others. It means also
the ability to separate
the target analyte
from all other components
(47)
(48)
(49)
Selectivity evaluation for pharmaceutical preparations
:
Identity of API should be tested, and its selectivity should be
proved.
Please be careful if the identification method that described
in Pharmacopeia/ Official method is not identical with its
in Pharmacopeia/ Official method is not identical with its
assay method (e.g. Identification: IR, color reactions; assay :
HPLC
)
Why for some cases, identification of API in pharmaceutical preparations
should be used by using > 1 methods: IR, TLC, Color reactions
(50)
“IR Library” should be validated routine
(51)
(52)
(53)
A B
C R & D/QC Lab. at Pharmaceutical Industry
Nice if related compounds, possible degradation products,
Densitograms (λλλλ = 260 nm) obtained from: (1) solution of standard mometasone furoate, (2) extract from excipients of laboratory-made cream, (3) extract of laboratory-made cream, (4) solution of nipagin, (5) solution of nipasol, (6) extract of commercial ointment-1, (7) extract of commercial lotion, (8) extract of commercial cream-1, (9) extract of commercial ointment-2 and (10) extract of commercial cream-2. Peak identities: (A) mometasone furoate, (B) nipagin and nipasol, (C) unknown.
Wulandari, L, Tan, KS., Indrayanto,G. (2003), J. Liq. Chromatogr. R & T, 26, 109-117
A Standard products, known impurities can be tested
(54)
The target peak must be proved regarding its identity and purity
Peaks should be well separated
(F.Melianita, J. Witha, S.Arifin, WF.Karina, G. Indrayanto (2009). J. Liq. Chromatogr. R & T, 32, 567-577)
(55)
Rs should be not less 1.75
(56)
IF Impurities, related compounds, possible Degradation products
were unavailable
Cited from: D. Widiretani, I. Luailia, G. Indrayanto,
(57)
If we have no impurities and/or
degradation products; we should do
stressed
forced experiments to prove the selectivity.
Identity and purity check
of peaks A and B must be performed
Cited from: D. Widiretani, I. Luailia, G. Indrayanto,
J. Planar Chromatogr. 26(2013) , 37-42
(58)
(59)
Determination of Xanthorrhizol in the rhizomes of Curcuma xanthorrhiza by GC FID
Xanthorrhizol ?
(F. Melinata & G. Indrayanto (2006), Unpublished work)
Should tried
3 different conditions For FID, RID, ELSD (= Blind methods)
(60)
Purity Check using PDA detector (HPLC) or Densitometer (TLC-Scanner)
Cited from: M. Yuwono & G. Indrayanto, Profile Drugs Excipients and Related Methodology, Vol. 32, Elsevier, 2005.
(61)
M.W. Dong, Modern HPLC
for Practicing Scientist, Wiley, 2006
The purity of the peak can be calculated by
comparing the MS of the u, a and d of the TIC peak.
(62)
Acetylsalicylic acid
200 nm
N=12804, Tf=1.20
Evaluation of the shape of acetylsalicylic acid
Evaluation of the shape of acetylsalicylic acid
chromatogram using 5 wavelengths
chromatogram using 5 wavelengths
0 5 10 15(min)
200 nm
220 nm
240 nm
260 nm
280 nm
N=12804, Tf=1.20
N=12970, Tf=1.20
N=12924, Tf=1.20
N=12887, Tf=1.20
N=12924, Tf=1.20
(63)
(64)
Analysis using LC MS/MS triple Quad
Cited from Agilent LCMSD triple Quad training, Singapore, May 2007
(65)
TA Sasaki, Chrom. Tech. July/August 2008
(66)
(67)
(68)
(69)
(Mw: 868.04)
M. Distl et al. (2009) Potato Research 52, 39-56
(Mw: 852.07)
(70)
The power of MRM
MRM is nicer compare to SIR or SIM
(71)
Why we need do the “
Why we need do the “identity test
identity test” and “
” and “
purity test
purity test” of the peaks for
” of the peaks for each
each samples for
samples for
quantitative analysis ?
quantitative analysis ?
•
If the peak is not the target compound,
the conclusion will be not correct.
•
If “impurity components” are included
in the target peak, the quantitative-results
in the target peak, the quantitative-results
will be wrong
•
By Using LC-MS/MS or GC-MS/MS
the target peak doesn’t need to be pure, it can
contain > 1 compounds which have same target
ion;
the Identity of the peaks can be differentiated by using
MS/MS spectrum (but ratio of qualifier to quantifier ions
should be in the acceptable range )
(72)
(73)
Linearity
•
The linearity of a method is its ability to provide measurement results that are directly proportional to the concentration of the analyte, or are directly proportional after mathematical transformation.• The linearity is usually documented as the ordinary least squares (OLS) curve, or simply as linear regression curve, of the measured responses (peak area or height) as a function of increasing analyte concentrations
•
Peak area
shall be preferred compared to peak heights for making the calibration curve• The linearity of the detector can be obtained by diluting of the analyte stock solution, whilst the linearity of the analytical method can be determined by making a series of concentration of the analyte from independent
preparations (weighing, spiking)
(G. Indrayanto, Suciati, M. Yuwono (2010), in Encyclopedia Chromatography 3rd Edition,Taylor &Francis, pp.2336-2339)
(74)
(75)
•
Analytical Methods Committee, Analyst (1988), 113: 1469-1471.
•
W. Horwitz, Referee (1995), December, 2.
•
J. van Loco; M. Elskens; C. Croux; H. Beernaert. Accred. Qual.
Assur (2002), 7: 281-285
‘Correlation coefficient (r) can not be used alone
for proving the linearity (
r close to 1 is not
a reliable indicator of linearity
)
Assur (2002), 7: 281-285
•
G. Indrayanto; M. Yuwono, in:J. Cazes (ed), Encyclopedia of
Chromatography (Supp.), Marcel Dekker, Inc, New York, NY
10016, 2003.
•
P.Araujo, J.Chromatogr.B. (2009) 877, 2224-2234
•
etc.
(76)
Evaluation of linearity/calibration curve
•
Relative process standard deviation (Vxo)- Funk’set al. 1992 (< 5 %); “sdv” (CATS from CAMAG) < 5
•Mandel’ test (should be OK)
•Residual test (should be OK)
•ANOVA-linearity testing (should be OK, p < 0.01)
•RSD of the Plot of response factor Vs. concentration (RSD < 2.5 – 5 %) et al. (Xp must be less than the lowest concentration of
•Xp value- Funk’s et al. (Xp must be less than the lowest concentration of the calibration curve)
• % Y intercept < 2
• “r” value (can not be used alone), even “r” is close to one
•Fisher Ratio (according to Araujo 2009)
•G value, Quality Coefficient (According to D Beer et al., 2007)
(77)
(78)
(79)
Some mathematical equations
for evaluation of linearity (Funk’s
et al
., 1992)
(
i i)
Y xo XO
bx
a
y
y
y
S
S
S
X
S
V
+
=
−
=
=
⋅
=
∑
ˆ
,
ˆ
where
%
100
(
i i)
Y xo XO
bx
a
y
y
y
S
S
S
X
S
V
+
=
−
=
=
⋅
=
∑
ˆ
,
ˆ
where
%
100
i YXO
y
a
bx
N
S
b
S
=
+
−
=
=
,
ˆ
2
where
Y iXO
y
a
bx
N
S
b
S
=
+
−
=
=
,
ˆ
2
where
(
)
(
)
0.05 p and 2 -N f for factor student t t , 1 1 1 . . Y with . 1 1 . . 2 2 2 2 P 2 2 = = = − = + + + = − + + =∑
i∑
i XX XX table Y XX P table XO P X N X Q Q X N t S a Q b Y Y N t S X(80)
(81)
The importance of the Xp’s Value for testing the linearity
Xp values must be < X1
Lowest Highest
(82)
(83)
Testing of Homogeneity (Homoscedasticity)
of the Calibration Curve
(84)
(85)
Proving Linearity according to Araujo (2009)
(86)
– The calibration spots should be independent of each other,
therefore spots should be applied from different, independently
weighed solutions – not by diluting from one and the same
stock solution.
Journal of Planar Chromatography 23 (2010) 3, 173–179 DOI: 10.1556/JPC.23.2010.3.1 173
0933-4173/$ 20.00 © Akadémiai Kiadó, Budapest
stock solution.
– As a consequence, calibration spots should be applied using
uniform volume of solutions, but different concentrations, to
avoid the different band-broadening effects of the different
amounts of solution per spot or band
(87)
Working range (WR) of a calibration curve can be defined as:
“the ratio of upper concentration (X
u)
and
lower concentration (X
l)”
In QC of Pharmaceutical Industry, WR = 1.5 – 2
Our recommendation: ± 20 % from target concentration ( with n = 4/5)
If WR > 10, It is recommended to use
“a weighting regression calibration curve”
(cited from: S. Kromidas (ed.) , HPLC Made to Measure, Wiley-VCH, Weinheim, 2006,pp. 105-115)
(88)
Requirement of assay using Single point Calibration
• Linear response function OK
• Negligible constant systemic error (CL is not significantly different to zero, or
• Negligible constant systemic error (CL is not significantly different to zero, or Intercept < 2 % from target compound )
• Homogeneity of variance
(J. Ermer & P. Nethercote, Validation in PharmaceuticaAnalysis, Wiley-VCH, 2015).
Confidence range of intercept:
Qxx
Xc
N
1
.
S
.
t
a
=
C
2 y 2) -N (P, table(89)
Detection limit (DL), Quantification limit (QL):
•
Impurities
•
Degradation products
•
Cleaning validation
•
Heavy metals
•
Heavy metals
•
Toxic metabolites
•
Pesticides
•
Waste products
•
Bio-analytical methods
(90)
Detection Limit & Quantitation Limit
Detection limit (DL) defined as the lowest
concentration of an analyte that can be detected under the analytical condition to be used, but cannot be measured quantitatively.
Quantitation limit (QL) is the lowest concentration that can be determined with acceptable accuracy and precision under the analytical conditions. Generally QL can be estimated as 3 times of DL.
DL and QL for instrumental (chromatographic)
analytical methods can be defined in the term of the signal-noise ratio (2-3/1 for DL and 10/1 for QL)
By constructing a linear regression of relative low concentrations of analyte, accurate value of DL can be calculated, in this case DL = Xp. We
recommend using of 5-10 relatively low
(91)
(92)
Determination of DL using method of Funk et al (DL = Xa/Xp): QL = 3DL; X2 and X1 (highest and lowest concentrations of standards) should be as low as possible!
(93)
(94)
(95)
(96)
(97)
(98)
Accuracy/ Trueness and Precision
Accuracy/ Trueness and Precision
(99)
Evaluation of Accuracy
Accuracy or trueness of an analytical method is given by the extend by which the value obtained deviates from the true value.
Firstly accuracy can be determined by analyzing a sample with known
concentration and comparing between the measured and true value.
The second approach is by comparing test results obtained from new method with results from the existing method that known to be accurate.
The third and fourth approaches are based on the percent recovery of known analyte spiked into blank matrices or products. The last technique is known as standard addition method.
For spiked samples into blank matrices, it is recommended to prepare sample in 5 different concentrations in the level of 80-120% of the target concentration.
For standard addition method, the spiking concentrations are 30-60 % of the label claimed
(Cited from : Indrayanto, G & Yuwono, M. (2003), in : Cazes, J. Ed. Encyclopedia of Chromatography (Marcel Dekker), Supplement
(100)
(1)
(2)
It is not only identification of selected markers (Mars > 3), but should be assayed, and
R must be in the specification range, or Cp/Cpk >1 or BETI < Toxic ?
(3)
Final Remark for (QC Herbal) Drugs
• For ensuring the quality of herbal derived drugs a
complete quality control was needed. The content of
selected markers and chromatographic metabolite profile must be both determined and evaluated.
• Authentic reference standards were essential for
determining markers in herbal drugs, whilst for chemical profiling botanical reference materials are chemical profiling botanical reference materials are needed.
• The concentrations of toxic metabolites(s), heavy metals (Hg, Pb, Cd, As), pesticides, mycotoxin
should be below their MPL.
• All methods that will used for QC purposes must be
properly validated before the methods can be routine
(4)
(5)
(6)