ERROR MESSAGE LIST

APPENDIX I ERROR MESSAGE LIST

Error Message Possible Cause Countermeasure

D2 LAMP!

The D 2 lamp has burnt out.

Replace the D 2 lamp.

WI LAMP!

The WI lamp has burnt out.

Replace the WI lamp.

Xe LAMP!

The Xe flash lamp fails to come on.

Retry and if this error message reappears, then contact our service office.

WAVELENGTH

Retry and if this error message INITIALIZE error

An abnormality has been detected in

wavelength initialization.

reappears, then contact our service office.

HARDWARE!

An abnormality has been detected in

Check the accessory device.

operation of an accessory device

If the cause of trouble is not

connected to the spectrophotometer.

located, then contact our service office.

SIGNAL! (sector/chopper)

An abnormality has been detected in

Retry and if this error message

rotation of the sector mirror.

reappear, then contact our service office.

POSITIONER!

For the cell positioner or 6-cell turret,

Check if the cell positioner or

an initialize error has been detected.

6-cell turret is normally mounted to the instrument.

SIGNAL error!

An abnormality has been detected in

Retry and if this error message

rotation of the sector mirror.

reappears, then contact our service office.

Calibrate!

Wavelength calibration has not been

Check that there is no sample in

accomplished properly.

the sample compartment and retry wavelength calibration.

ACC OPERATION error!

An abnormality has been detected in

Check the accessory device.

operation of an accessory device

If the cause of trouble is not

connected to the spectrophotometer.

located, then contact our service office.

MOMENTARY POWER

Restart the spectrophotometer FAILURE!

The spectrophotometer main unit

has been powered off.

main unit and carry out initialization.

Attenuate!

Attenuation rate measurement has

Carry out attenuation rate

not been executed.

measurement.

Battery!

The backup battery voltage has

The backup battery needs to be

dropped.

replaced. Contact our service office.

WAVELENGTH DRIVE!

An abnormality has been detected in

Retry and if this error message

the wavelength drive system.

reappear, then contact our service office.

CAPACITOR CAPACITY!

An abnormality has been detected in

Retry and if this error message

setting of the capacitor.

reappears, then contact our service office.

DATA FILE SUM error

An abnormality has been detected in

Retry and if this error message

the Backup RAM.

reappears, then contact our service office.

METHOD FILE SUM error

An abnormality has been detected in

Retry and if this error message

the Backup RAM.

reappears, then contact our service office.

Error Message Possible Cause Countermeasure

BACKUP RAM SUM error

An abnormality has been detected in

Retry and if this error message

the Backup RAM.

reappears, then contact our service office.

BATTERY VOLTAGE

Retry and if this error message DROP

The backup battery voltage has

dropped.

reappears, then contact our service office.

ROM STATUS error

An abnormality has been detected in

Retry and if this error message

the ROM.

reappears, then contact our service office.

RAM STATUS error

An abnormality has been detected in

Retry and if this error message

the RAM.

reappears, then contact our service office.

sum error

A sum error has been detected.

Retry and if this error message reappears, then contact our service office.

WAVELENGTH

Retry and if this error message CALIBRATION SUM error

An abnormality has been detected in

wavelength calibration or correction

reappears, then contact our

data.

service office.

SYSTEM DATA SUM error

An abnormality has been detected in

Retry and if this error message

system data of the instrument.

reappears, then contact our service office.

LAMP ON DATA SUM

Retry and if this error message error

An abnormality has been detected in

lamp turn-on data.

reappears, then contact our service office.

SERIAL NO. SUM error

An abnormality has been detected in

Retry and if this error message

serial no. data.

reappears, then contact our service office.

Retry and if this error message SUM error

PARAMETER INITIALIZE

An abnormality has been detected in

data on parameter initialization.

reappears, then contact our service office.

COMMAND SIGNAL error

Retry and if this error message (failure in sector & chopper

An abnormality has been detected in

reappears, then contact our rotation mechanism)

rotation of the sector mirror.

service office.

VIS/NIR CAM CHANGE

Retry and if this error message INITIALIZE error

An abnormality has been detected in

changeover between the visible and

reappears, then contact our

near infrared regions.

service office.

LIGHT INTENSITY

Retry and if this error message CONTROL MOTOR

An abnormality has been detected in

reappears, then contact our INITIALIZE error

the light intensity control motor.

service office.

SLIT INITIALIZE error

The slit initialization position cannot

Retry and if this error message

be detected.

reappears, then contact our service office.

FILTER INITIALIZE error

The filter initialization position cannot

Retry and if this error message

be detected.

reappears, then contact our service office.

EEPROM SUM error (1)

An abnormality has been detected in

The default value is set.

wavelength calibration value data,

Execute the following.

RS reversal state data or HV zero adjustment correction value data

(1) Wavelength calibration

stored in the EEPROM.

(2) R/S reverse (3) Detector zero correction

Error Message Possible Cause Countermeasure

EEPROM SUM error (2)

An abnormality has been detected in

The default value is set.

system baseline data stored in the

Execute system baseline

EEPROM.

measurement.

EEPROM SUM error (3)

An abnormality has been detected in

“0” is set.

lamp turn-on time data stored in the

The previous value is recorded in

EEPROM.

the event log. For control, use this time data for summation.

EEPROM SUM error (4)

An abnormality has been detected in

The default value is set.

serial number data or

Contact our service office.

spectrophotometer type data stored in the EEPROM.

MAIN CPU ROM STATUS

Retry and if this error message error

An error has occurred in the main

CPU ROM.

reappears, then contact our service office.

MAIN CPU RAM STATUS

Retry and if this error message error

An error has occurred in the main

CPU RAM.

reappears, then contact our service office.

SUB CPU ROM STATUS

Contact our service office. error

An error has occurred in the sub

CPU ROM.

SUB CPU RAM STATUS

Contact our service office. error

An error has occurred in the sub

CPU RAM.

SUB CPU VERSION

Contact our service office. inconsistent

The sub CPU program version is

inconsistent with the main CPU program version.

WL MOTOR CONTROL

Contact our service office. error

An error has been detected in

wavelength motor control.

SECTOR MOTOR control

Contact our service office. error

An error has been detected in sector

motor control.