53
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
26 Must have Supporting letter from principal and supporting letter from local support
Distributor in indonesia
f. RF Accessories and cabling
1 3.5 mm to Type-N adapter kit
2 Type-N accessory kit, 50 Ohm
3 Dua type-N m to BNC f dan Dua type-N m to BNC m adapters
4 Dua type-N f to BNC f dan dua type-N f to BNC f adapters
5 BNC male short
6 RF Cabling Type-Nm to Type-Nm, 50 ohm nominal impedance
3.28 ft. or 1 meter 7
Max frequency: 18 GHz 8
Typical VSWR: 1.19 9
Typical insertion loss: 1.13 dB 10
Typical phase stability: ± 2.0 degrees 11
Typical amplitude stability: ± 0.05 dB 12
Nominal velocity of propagation: 85, Nominal dielectric constant: 1.4 13
RH799: 70 to 1000 MHz wide-band antenna 14
Length: 113.5 cm full length, 19.5 cm retracted 10 stages 15
Weight: 65 g 16
Gain: 2.15 dBi 17
Maximum power rating: 10 W FM 18
Impedance: 50 ohms 19
Type: ¼ wave 70 to 300 MHz, 180 degree tilt angle adjustable, telescopic whip antenna
20 Connector type 3.5 mm m-m, f-f, m-f
21 Frequency range DC to 26.5 GHz
22 Typical min return loss -32dB
23 Typical max insertion loss 0.074 dB
24 Must have Supporting letter from principal and supporting letter from local support
Distributor in indonesia
g. Device analyzer and generator voltage power
1 SourceMeasure Unit for Battery Drain Analysis, 2-Quadrant, 20V1A or 6V3A, 20W
2 DC Power Module, 2-Quadrant
3 Battery ChargeDischarge Module, 0 to 8V -2 to 3A, 24W
4 DC Power Module 60V, 5A, 300W
5 DC Power Analyzer Mainframe
6 slot mainframe holds up to 600 W of total power and up to 4 modules.
7 Voltmeter accuracy: Up to 0.025 + 50 µV, up to 18 bits
8 Ammeter accuracy: Up to 0.025 + 8 nA, up to 18 bits
9 Arbitrary waveform generator function: Bandwidth up to 100 kHz, output power up to
500 W 10
Scope function: Digitizes voltage and current at up to 200 kHz, 512 kpts, up to 18 bits 11
SourceMeasure Unit 12
Output Ratings: Voltage: Up to 20 V
13 Current: Up to ±3 A
14 Power: 20 W
15 Battery Drain Analysis, 2-Quadrant
16 Programming Accuracyat 23°C ± 5°C after 30 minute warm-up. Applies from min. to
max. program:
54
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
17 Voltage, 20 V range: 0.025 + 1.8 mV
18 Voltage, 6 V range: 0.025 + 600 µV
19 Current, 3 A and 1 A ranges: 0.04 + 300 µA
20 Current, 300 mA range: 0.03 + 150 µA
21 Transient Response: 30 mV of droop in response to 0.15A µs current transients
22 Measurement Accuracy at 23°C ± 5°C:
Voltage, 20 V range: 0.025 + 1.2 mV 23
Current, 3 A range: 0.03 + 250 µA 24
Current, 100 mA range: 0.025 + 10 µA 25
Current, 10 µA range: 0.025 + 8 nA 26
DC Power Module, 2-Quadrant 27
Output Ratings: Voltage: 8 V
28 Current: -2 A to +3 A
29 Power: 24 W
30 Programming Accuracy• Voltage: 0.1 + 10 mV
31 Positive Current: 0.1 + 1.8 mA
32 Negative Current -2 A: 0.2 + 1.8 mA
33 Measurement Accuracy at 23°C ± 5°C:
Voltage: 0.05 + 5 mV 34
Current high range: 0.1 + 600 µA 35
Current low range ≤ 150 mA: 0.1 + 75 µA 36
DC Power Module 37
Output Ratings: Voltage: 60 V, Current: 5 A, Power: 300 W
38 Programming AccuracyVoltage: 0.1 + 60 mV Current: 0.15 + 60 mA
39 Ripple Noise 20 Hz to 20 MHz
Voltage peak-to-peak: 35 mV, Voltage rms: 9 mV, Current rms: 6 mA
40 Support 4-slot mainframe holds up to 600 W of total power and up to 4 modules
modular dc power supply 41
measure: DMM, Votlt , Ampere ,oscilloscope , data logger. Arbitrary waveform generator.
42 Ammeter accuracy: Up to 0.025 + 8 nA, up to 18 bits
43 Arbitrary waveform generator function: Bandwidth up to 100 kHz, output power up to
500 W 44
Scope function: Digitizes voltage and current at up to 200 kHz, 512 kpts, up to 18 bits 45
Data logger function: Measurement interval from 20 µs to 60 s, max of 500 M readings per datalog
46 GB of non-volatile data storage for data log, scope traces, instrument settings
47 Heightwidthdepth 194.7 mm425.6 mm313 mm
48 Weight with 4 modules typical 16 kg
49 AC input
50 Input ratings ~ 100 VAC – 240 VAC; 5060400 Hz
51 Power consumption 1440 VA
52 Power factor 0.99 nominal input and rated power
53 Must have Supporting letter from principal and supporting letter from local support
Distributor in indonesia
h. Measuring device of signal strength