Specific measuring device for motor inspection

84 SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011. j Test voltage dc : 250V, 500V, 1000V, 2500V, 5000V, 10000V k Accuracy : 5 for 200 kΩ to 200 GΩ l Bar graph range : 0 to 1 TΩ m Insulation test voltage accuracy : -0 , +10 at 1 mA load current n Induced ac mains current rejection : 2 mA maximum o Charging rate for capacitive load : 5 seconds per μ F p Leakage current measurement : 1 nA to 2 mA Accuracy: ± 5 + 2 nA q Capacitance measurement : Range: 0.01 uF to 15.00 μ F Accuracy: ± 15 rdg + 0.03 μ F r Live circuit warning : Warning range: 30 V to 660 V acdc, 5060 Hz Voltage Accuracy: ± 15 + 2 V 8 Must have Supporting letter from principal and supporting letter from local support Distributor in indonesia

b. Specific measuring device for motor inspection

1 Motor and drive inspection kit using insulation tester combine with true-rms digital multimeter in one compact tool, which provides maximum versatility for both troubleshooting and preventative maintenance. Used on motors, generators, cables, switch gear, etc. Current clamp is provided to measure 1A AC to 400A AC with easy reading. Including phase rotation indicator or easy determining field rotation and phase sequence, and soft case. Other small current clamp able to measure leakage current with 1µA resolution. Measuring range : Max 1000V DC, accuracy ±0.09+2, max resolution 0.001V ;Max 1000V AC, accuracy ±2+3, max resolution 0.1 mV ; Max 400mA DC, accuracy ±0.2+2, max resolution 0.01 mA ; max 400mA AC, accuracy ±1.5+2, max resolution 0.1 mV ; Max 50 MΩ , accuracy ±0.9+2, max resolution 0.1 Ω ; Max 9,999µF, accuracy ±1.2+2, max resolution 1nF ; Max 100kHz, accuracy ±0.1+1, max resolution 0.01 Hz ; Max 537°C, for temperature measurement ; Range 6V, resolution 1mV, accuracy±2+1 for diode testTest voltage : 50, 100, 250, 500, 1000 V Maximumresistance : 2 GΩ Insulation resolution : 0.01 MΩ for test voltage at 50V and 100V Accuracy : ±1.5 + 5 for test voltage 500V and 1000V Overvoltage category : ANSIISA 82.02.01 61010-1 2004, CANCSA-C22.2 NO. 61010-1-04, and IECEN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 VSpecification for current clamp : 1A to 400A, accuracy 2 + 0.06 A, max voltage 1000V AC rms from earth to ground, in compliance with EN61010 ; max conductor diameter 32 mm, max conductor size 750 MCM, shrouded banana plugs Safety rating for current clamp : CAT IV 600 V CAT III 1000 V Specification for phase indicator : 40-70V, 15-400 Hz, continuous operating time, 3-phase indication via LCD, no battery required Specification for leakage current meter : Digital LCD display 3200 counts, auto-range, range 3 mA, 30 mA, 30 A, 60 A, current resolution 1µA, 0.01mA, 0.01A, 0.1A, accuracy up to 1+5, auto power off, 40 mm conductor diameter max, measurement cycle 2 timessecond Digital display, 12 timessecond Bar-graph display High current clamp-on amp meter measurement up to 2000A, with wide jaw, accurate peak measurements can be taken using the in-rush current mode, for motors and inductive loads. Also can measure voltage and resistance, use the low-pass filter to smooth out noisy loads.Measurements : AC current, AC voltage, resistance, continuity, DC volts, DC current, true-rms, frequency, AC+DC voltage, AC+DC current, minmaxavgCurrent measurement dc : Range 40A ; resolution 10mA, accuracy 1.5 rdg + 15 digits ; Range 400A ; resolution 100mA, accuracy 1.5 rdg + 5 digits ; Range 2000A ; 1400 ac rms, accuracy 1.5 rdg + 5 digits. 85 SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011. 2 Current measurement ac : Range 40A ; resolution 10mA, accuracy 10A : 3.5 rdg + 15digits ; Range 400A ; resolution 100mA, accuracy 10A : 3.5 rdg + 5digits ; Range 2000A ; 1400 ac rms, accuracy 10 A : 3.5 rdg + 5digits 3 Crest factor : 2 1000 A, 2.4 833 A for range 2000A ; 1400 ac rms 4 Voltage measurement ac and dc : Range 4 V, 40 V, 400 V, 600 V ac rms, 1000 Vdc ; Accuracy 1 rdg + 10 digits for dc voltage ; Accuracy 3 rdg + 5 digits for ac voltage 5 Ohms measurement : Range 400Ω , 4kΩ , 40kΩ , 400kΩ ; accuracy 1.5 + 5 digits 6 Frequency measurement : 5.0 Hz to 1 kHz, resolution 0.1 Hz 15 Hz to 399.9 Hz; 1 Hz 400 Hz to 1 kHz, accuracy up to 0.2 + 2 counts 7 Other features : Large 58 mm jaw, high voltage measurement of 1000 ac + dc true-rms, 600V ac and 1000V dc, inrush current mode, low-pass filterRugged compact oscilloscope with true-rms digital multimeter and paperless recorder, that also can convert waveform information into a harmonics display using FFT processing which shows the relative amplitudes of the 1st up to the 33rd harmonic. Including BNC adapter, battery charger, shielded test lead sets, hook clips, ground leads, hard carrying case, and software with adaptercable. 8 Bandwidth : 40 MHz 9 Input meter : Dual input Vdc, Vac, Vac+Vdc, Ohms,Continuity , Diode-test Current, °C, °F, Capacitance, dBV, dBM, Crest Factor,Touch Hold and Zeroset 10 Cursors : ∆T, 1∆T, V, ∆V, rise- and fall time 11 Trendplot recording : Automatic vertical scaling and time compression, display the actual, MIN, MAX and AVG reading 12 Real-time sample rate : 25 MSs 13 Equivalent sampling rate : up to 2.5 GSs 14 Timebase range : 20 ns - 1 mindiv 15 Trigger types : Connect-and-View™, Free Run, Single Shot, Edge, Video, External 16 Glitch capture : 40 ns 17 Other features : Measure Watt, VA, VAR, Power Factor PF with Single phase or balanced three-phase delta-configuration mains supply. Has trendplot, cursors, industrial bushealth measurement gives indication for electrical signals on industrial buses and networks, harmonics THD in r or f, waveform mathematics Voltage waveformCh.A, Current wavefrom Ch.B or Power Ch.A x Ch.B, automatically generated 18 Battery life : 7 hrs using NiMH 19 Electrical Safety : 600V CAT III EN61010-1 20 Non-contact thermometer with visual and thermal camera, has hot and cold trackers, free reporting software, automated monitoring alarm, including hardcase, micro-SD card, rechargeable battery with charging base. 21 Measurement range : -10 °C to +250 °C ; Measurement accuracy : ±2 °C or ±2 22 Field of view : 28° x 28° 23 Storage medium : Store up to 10,000 images per Gb 4Gb micro SD card included 24 Image optic system : PyroBlend™ Plus optic 25 Battery type : Li-Ion rechargeable 26 File formats : .is2 format saved to SD card, but can export to BMP, DIB, GIF, SPE, FIF, SPEG, JPG, PNG, TIF and TIFF 27 Other features : Built-in digital camera, thermal heat map overlay with 5 blending modes, ergonomics, focus free 28 Safety and compliance : CFR47: 2009 Class A. Part 15 subpart B; CE: EN 61326:2006; IECEN 61010-1:2010 29 Vibration screening device with severity-scaled readings of overall vibration and bearing condition, also can measure temperature in one tool, exportable data via USB, trending 86 SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011. in Microsoft Excel using built-in templates, on-board memory up to 3,500 measurements, audio output for listening to bearing tones directly, external accelerometer support for hard to reach locations, has flashlight and large screen. Provides Crest Factor + for bearing assessment. Include tool bag for motor inspection kit. 30 Low frequency range : 10 Hz to 1,000 Hz 31 High frequency range : 4,000 Hz to 20,000 Hz 32 Severity levels : Good, Satisfactory, Unsatisfactory, Unacceptable 33 Vibration meter limit : 50 g peak 100 g peak-peak 34 Sampling rate : 20,000 Hz low frequency, 80,000 Hz high frequency 35 Signal to noise ratio : 80 dB 36 Sensor sensitivity : 100 mV g ± 10 37 Sensor measurement range : 0.01 g to 50 g, resolution 0.01 g 38 Amplitude units : g, msec² acceleration, insec, mmsec velocity, mils, mm displacement 39 Real time clock backup : coin battery 40 Temperature measurement : -20 °C to 200 °C , accuracy ±2 °C 41 Connectors : USB mini-B 7-pin, stereo audio output jack 3.5 mm Audio Plug, external sensor jack SMB connector 42 Must have Supporting letter from principal and supporting letter from local support Distributor in indonesia

c. Specific measuring device for area compliance inspection