56
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
22 300 V switching
23 Dual 4-channel RF multiplexter 2 GHz bandwidth
24 Multifunction module 16 bits of digital input and output
25 100 kHz totalizer input
26 Two ±12 V analog output
27 Included
Kelvin clip, BenchLink DataLogger, Thermocouple kit 10 x J-type thermocouples,
28 10kOhm thermistors 0 to 75 degree C, Cable Kit
29 Support LAN USB IO to PC
30 Support USB flash drive untuk penyimapanan dan transfer file.
31 Include free Graphical Web Interface
32 Must have Supporting letter from principal and supporting letter from local support
Distributor in indonesia
k. Low frequency modulation signal generator
1 Wave Form
Standard : sine, square, pulse,ramp, noise, sinxx, exponential rise, exponential fall, cardiac, DC volts
2 Arbitrary : Waveform length 1 to 64K points
3 Amplitude resolution 12 bits including sign
4 Repetition rate 1 µHz to 25 MHz
5 Sample rate 200 MSas
6 Filter bandwidth 50 MHz
7 Non-vol. memory Four 4 64K wave- forms
8 Output Amplitude : into 50Ω 10 mVpp to 10 Vpp
9 Accuracy : at 1 kHz, 10 mVpp, Autorange on± 1 of setting ± 1mVpp
10 Flatness : sinewave relative to 1 kHz,Autorange on
10 MHz ± 1 0.1 dB 10 MHz to 50 MHz ± 2 0.2 dB
50 MHz to 80 MHz ± 5 0.4 dB 11
Units : Vpp, Vrms, dBm,high and low level 12
Resolution : 0.1 mV or 4 digits 13
Offset : into 50Ω ± 5 Vpk ac + dc 14
Accuracy : 1 of setting + 2 mV + 0.5 of amplitude 15
Waveform Output :Impedance 50Ω typical fixed 10 MΩ output disabled 16
Isolation : 42 Vpk maximum to earth 17
Protection : short-circuit protected, overload relay,automatically disables main output 18
Included Certificate of calibration - ANSINCSL Z540, printed, Transit Case,
Accessory Pouch 19
Color display with graph mode for visual verification of signal settings 20
Built-in multiple-unit link for synchronous channels 21
Must have Supporting letter from principal and supporting letter from local support Distributor in indonesia
l. System for measuring device and analyzer for Digital Radio Testing
1 IQ Gen Modulation for IQ Creator waveforms
2 Audio Analyzer, 1 MHz - 1 GHz Operation
3 Auto-Test II Analog
4 Harmonics Spurious Measurements
5 Tracking Generator
6 Analog Duplex Occupied Bandwidth and Power Between Markers
7 P25 Conventional Operation with DES OFB Type III
57
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
8 P25 Trunking Operation VHF UHF 700 800MHz
9 SmartnetSmartzone Option
10 KVL Keyloader
11 Analog Simulcast Option
12 Autotest II for P25 Radio Systems
13 P25 AES Encryption
14 Occupied Bandwidth for P25
15 Motorola ASTRO 25 series Autotest Alignment Software
16 Motorola ASTRO series Autotest Alignment Software
17 Motorola ASTRO 25 series XTL Power Autotest Alignment Software
18 TIAEIA-603 FM Land Mobile Test Software
19 Motorola APX™ series Autotest Alignment Software
20 DMM Test Leads
21 10 AMP Current Shunt 0.01 Ohm.
22 Scope Probe Kit
23 Microphone
24 TETRA Mobile Station Test
25 TETRA Base Station Test
26 TETRA Direct Mode Test
27 TETRA Energy Economy Mode
28 TETRA Autotest II
29 DMR MOTOTRBO™ ETSI 102-361
30 Autotest II for DMR Radio Systems
31 DMR XML Channel Logger Option
32 dPMR - ETSI 102-658
33 Autotest II for dPMR Radio Systems
34 dPMR Control Channel Logger
35 NXDN
36 Autotest II for NXDN Radio Systems
37 TETRA Cassidian BS3 Base Station Test Software
38 DMR Repeater Autotest Software
39 Kenwood NXDN series Autotest Alignment Software
40 RF Signal Generator
a Frequency Range
: 10 MHz to 1.05 GHz Standard Usable from 100 kHz
b Resolution
: 1 Hz c
Accuracy : Frequency standard ±1 count
d Output Level Range
: TR Port: -130.0 to -30.0 dBm -30 dBm max for CW or FM; -35 dBm max for AM modulations; -40 dBm max for complex modulations :
Duplex: -130.0 to +10.0 dBm +10 dBm max for CW or FM; +5 dBm max for AM modulations; 0 dBm max for complex modulation
e Resolution : 0.1 dB
f Accuracy : 1.0 dB for levels -110 dBm Typical better than 0.6 dB
: 1.5 dB for levels ≤-110 dBm Typical better than 1.0 dB g
Spectral Purity 1
Residual FM : 5 Hz 300 Hz to 3 kHz bandwidth
2 Residual AM
: 0.1 RMS 300 Hz to 3 kHz Bandwidth 3
Harmonics : -25 dBc Typically -30 dBc, RF level set at
+10 dBm 4
Non-Harmonics : -55 dBc all freq. except Crossovers,
58
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
: -35 dBc At 2nd order crossover frequency 10 MHz to 1 GHz: Crossover = 1400 MHz - Gen freq.
Phase Noise : -110 dBcHz 10 kHz offset, RF 500 MHz
: -106 dBcHz 10 kHz offset, RF ≤1000 MHz : -95 dBcHz 10 kHz offset, RF 1000 MHz
h Modulation Selections : OFF, AM, FM, FM50us, FM75us, FM750us, AM USB,
AM LSB, IQGEN i
Waveforms : Sine, Square, Triangle, Ramp, DCS,DTMF
j Internal FMDeviation Range
: ±0.001 to ±150 kHz, OFF k
Accuracy : 3 From ±1 kHz to ±100 kHz deviation, 20 Hz to 15 kHz rate
l Internal AM
1 Modulation Range: 0 to 100
2 Accuracy : 1Modulation from 10 to 90 20 Hz to 15 kHz rate
m Internal SSB
1 Modulation Selection
: Upper SideBand USB or Lower SideBand LSB
2 Resolution
: 0.1 n
Eksternal AMFMSSB 1
Audio input : With 1 Vrms, AMFMSSB have same characteristics
as internal sources, ±10 of indicated setting. Audio 1 or Audio 2 input from 20 Hz to 15 kHz [300 Hz to 3 kHz SSB] unbalanced. 8 Vrms
maximum modulation input level.
2 Microphone input : With 50 mVrms, AMFMSSB have same
characteristics as internal sources, ±10 of indicated setting. MIC Input from 100 Hz to 15 kHz [300 Hz to 3 kHz SSB].
o Internal IQ Gen
1 Sample Rate
: 1.89 Msamplessec 2
Size : 3.8 million samples
41 RF Receiver
a Demod Selections : AM, FM, FM50us, FM75us, FM750us, AM USB, AM LSB
b Sensitivity : -100 dBm 10 dB SINAD, FM, 25 kHz, 1 kHz rate, 6 kHz FM
Deviation, 300 Hz to 3.4 kHz AF Filter, Pre-amp OFF c
Sensitivity : -113 dBm 10 dB SINAD, FM, 25 kHz, 1 kHz rate, 6 kHz FM Deviation, 300 Hz to 3.4 kHz AF Filter, Pre-amp ON
d Demod Output Level
e FM
: Nominally 1 Vrms for deviation ±14 of selected BW; 25 kHz BW same output level as 30 kHz BW
f AM
: Nominally 2 Vrms 100 AM 42
RF Measurements a
RF Power Meter Broadband b
Level Range : 100 mW to 125 W Usable from 10 mW
c Resolution
: 4 digits for W or 0.1 dB d
Signal : CW, FM, C4FM, 4FSK 43
RF Power Meter Inband a
Level Range : TR Port: -60 to +51 dBm
b Lowest reading is receiver BW dependent Narrower bandwidths can measure
lower levels.
59
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
c ANT Port: -100 to +10 dBm
d Lowest reading is receiver BW dependent Narrower bandwidths can measure
lower levels. e
Resolution : 0.1 dB
f Signal : CW, FM, AM, C4FM, 4FSK, QPSK, QAM
44 RF Counter
a Level Range for Auto-tune : TR Port: -10 to +50 dBm Find level is selectable
ANT Port: -60 to +10 dBm Find level is selectable b
Signal : CW, FM, AM 70 modulation c
RF Error Meter d
Level Range : TR Port: -10 to +50 dBm
ANT Port: -60 to +10 dBm e
Signal : CW, FM, AM 70 modulation f
Demodulation Measurements g
RF Characteristics h
Input RF Level : TR Port: -10 to +50 dBm ANT Port: -80 to +10 dBm
45 Demod Counter
a Range : 20 Hz to 20 kHz 1 to 100 kHz FM Deviation, IF BW set appropriately
for the received modulation BW b
20 Hz to 10 kHz 30 to 90 AM, IF BW set appropriately for the received modulation BW
c Accuracy
: ±50 ppm ±10 ppm typical 46
FM Deviation meter a
Range : 0 to 150 kHz b
FM Rate: 20 Hz to 20 kHz IF BW set appropriately for the received modulation BW
47 AM Deviation Meter
a Range : 0 to 100
b AM Rate
: 20 Hz to 15 kHz IF BW set appropriately for the received modulation BW
48 SINAD meter
a Range : 0 to 60 dB
b Accuracy
: ±1 dB, ±1 count SINAD 3 dB, ≤40 dB, 5 kHz LP AF filter c
Frequency Range : 300 Hz to 5 kHz
d Level Range Audio
: 0.1 to 30 Vrms e
Distortion meter Resolution
: 0.1 f
Accuracy : ±0.5 Distortion 1 to 10, 5 kHz LP AF Filter
g HUM and Noise
Range : -100 dB to 0 dB h
Resolution : 0.01 dB
i Accuracy
: ±1 dB, ±1 count -60 dB, ≤-20 dB j
Signal To Noise Ratio Range : -100 to 0 dB
k Resolution
: 0.01 dB l
Accuracy : ±1 dB, ±1 count -60 dB, ≤-20 dB
49 Audio Function Generator
a Wave Shape
: Sine, Square, Triangle, Ramp, Digital Coded Squelch, DTMF b
Frequency range : Sine: 20 Hz to 40 kHz usable 1 Hz to 40 kHz
c Square, Triangle and Ramp: 20 Hz to 4 kHz usable 1 Hz to 40 kHz
60
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
d Level range
: 1 mV to 5V RMS into a 10 kW load 50
Oscilloscope Trace Types : Live, captured, accumulated
Marker Functions : Time with amplitude, deviation or depth
Delta marker including 1Δ t, e.g. Hz Digital Multimeter
ACDC Voltmeter Full Scale Range
: 200 mV, 2 V, 20 V, 200 V, 2000 V, Auto ACDC Ammeter
Full Scale Range : 200 mA, 2 A, 20 A, Auto 20 A range uses optional shunt
connected to Voltmeter 51
Ohmmeter Full Scale Range
: 200 ohms, 2 kohms, 20 kohms, 200 kohms, 2 Mohms, 20 Mohms, Auto
52 RF Spectrum Analyzer
a Freq range
: 10 MHz to 1.05 GHz standard Usable from 100 kHz b
Span mode : StartStop, CenterSpan and Zero Span
c Level
d Ref Level Range: TR Port: -50 to +50 dBm
: ANT Port: -90 to +10 dBm e
Vertical Scales : 1, 2, 5, 10 dBdivision f
Dynamic Range : 70 dB Antenna, no attenuation, Ref Level -30 dBm, 30 kHz RBW
g Accuracy
: ±1 dB Input signal -10 dB from Ref Level, Normalized,preamp off
h Harmonic Spurious
: -55 dBc Input Level of -30 dBm, Ref Level at -20 dBm
i Non-Harmonic Spurious : -60 dBc Input Level of -30 dBm, Ref Level at -20
dBm j
Displayed Average Noise Level DANL : -125 dBm Typical, 300 Hz RBW, ANT Port terminated, 20 sweep average
k Resolution Bandwidth
l RBW Selections : 300 Hz, 3 kHz, 30 kHz, 60 kHz, 300 kHz, 6 MHz
m RBW 60 dB3 Filter Shape
: 10:1 n
Selectivity - Filter Shape : 60 dB3 dB ratio better than 10:1 o
Video bandwidth Range : 10 Hz to 1 MHz in a 1, 3, 10 sequence, plus NONE
p Sweep
Frequency Sweep Time : 100 mS to 100 S in a 1, 2, 5 sequence Zero Span Sweep Time : 50 mS to 100 S in a 1, 2, 5 sequence
Sweep Trigger Source : Internal and external
q Marker
Track : Frequencies or time and amplitudes
Number of Markers : 8
Marker Functions : Marker to Peak
Marker to Next RightLeft Marker to Minimum
Marker to Ref Level Marker to Center Frequency
Marker sets Span, Marker sets Vertical Scale Zero Span only
61
SPEKTEK ALAT LABORATORIUM PENGEMBANGAN INDUSTRI TEKNOLOGI KEPOLISIAN KSA T.A. 2011.
53 Must have Supporting letter from principal and supporting letter from local support
Distributor in indonesia
m. Audio Analyzer measuring device